Joerns® Assist Device and Side Rails

Vari-Care Models

© 2014 Joerns Healthcare • 6110015 RevJ • 14-2315

Joerns Side Rails/Assist Devices should be 

assembled, installed, utilized and maintained in the 

manner described in their accompanying Joerns User/ 

Service Manual. The instructions for use of these 

devices also list the bed type or model with which they 

may be used.

This product is only one part of your healthcare bed 

system. Proper combinations of mattress, head/foot 

panels and assist devices are needed to minimize 

the risk of entrapment. For more information, call 

Customer Care, 800-826-0270.

Save These Instructions

Thank you and congratulations for choosing Joerns 

Vari-Care assist devices and side rails. Joerns Assist 

Devices are developed with safety, comfort and 

convenience of both residents/patients and caregivers 

in mind. 

Advanced ergonomic design will provide years of 

reliable service in your facility.

Bed System Entrapment Information

Although essential in the practice of long-term care, 

bedside rails, in recent years, have also been a 

subject of regulatory review and evolution in design 

and use.

That focus includes not only the challenge of achieving 

an appropriate balance between resident security and 

unnecessary restraint, but also the additional safety 

issue of entrapment.

The U.S. Food and Drug Administration (FDA), 

working with our company and other industry 

representatives has addressed the potential danger 

of entrapment with new safety guidelines for medical 

beds. These guidelines recommend dimensional limits 

for critical gaps and spaces between bed system 

components.

Entrapment zones involve the relationship of 

components often directly assembled by the 

healthcare facility rather than the manufacturer. 

Therefore, compliance is the responsibility of the 

facility.

As the leading manufacturer of long-term care beds 

and a frontrunner in addressing this critical issue, 

Joerns Healthcare can offer you the expertise, 

assistance and products to bring your facility into 

compliance.

Joerns

®

 Compliance Solutions

Matching the right bed components in order to meet 

regulatory guidelines can be complex.

That is why Joerns offers a wide array of compliance 

options. We assist customers in selecting compliant 

accessories recommended for their specific bed 

model.

Creating a Safer Care Environment

While the guidelines apply to all healthcare settings, 

(hospitals, nursing homes and at home), long-term 

care facilities have particular exposure since serious 

entrapment events typically involve frail, elderly or 

dementia patients.

For More Information

To learn more about compliance options with Joerns 

products, visit our website at www.joerns.com, or 

contact our Customer Care reps at 800-826-0270 and 

ask for free informational publications.

To learn more about entrapment zones, assessment 

methods and guidelines concerning entrapment, 

contact Joerns Healthcare at 800-826-0270 or 

consult the FDA website:  http://www.fda.gov/

MedicalDevices/ProductsandMedicalProcedures/

GeneralHospitalDevicesandSupplies/HospitalBeds/

default.htm. 

Table of Contents

Important Precautions .................................................................................................................2
Entrapment Information ...............................................................................................................3

Assist Device/Side Rail Options

 ..................................................................................................4

Bed Compatibility.........................................................................................................................5
Installation Instructions  ...............................................................................................................6
Warranty ....................................................................................................................................12

Summary of Contents for Vari-Care F14SC

Page 1: ...User Service Manual Joerns Assist Device and Side Rails Vari Care Models To avoid injury read user s manual before using...

Page 2: ...lowed Warning An optimal bed system assessment should be conducted for each resident by a qualified clinician or medical provider to ensure maximum safety of the resident The assessment should be cond...

Page 3: ...lines recommend dimensional limits for critical gaps and spaces between bed system components Entrapment zones involve the relationship of components often directly assembled by the healthcare facilit...

Page 4: ...6110015 RevJ 14 2315 Matching Side Rail Options to your Beds F16C F168 F1576C for 76 beds 58 00 15 00 11 25 IN RAIL BODY 4 41 LARGEST GAP OF SLEEP DECK APPROX 4 30 TO TOP 67 00 11 25 IN RAIL BODY 4 4...

Page 5: ...Combination For older model beds contact product service Joerns Healthcare recommends the use of a support surface with minimum dimensions of 35 in width by 6 in depth Support surfaces used in conjunc...

Page 6: ...he bed F14SC For B530 B330 and EasyCare 2000 2004 Note Latch must face the center of the bed F1580C F1576C F16C for 80 EasyCare 2000 2004 Note Latch must face the foot end of the bed F1576C For 76 Eas...

Page 7: ...w 4 6 F168 Not included in N832 Full Rail Mount Kit Shown for reference only Suggested tools 2 wrench or socket Installation Instructions 1 Raise the back and foot sections of the bed as shown in Figu...

Page 8: ...ustrations When complete there should be an alignment label on each side of the frame aligned with the cross member located closest to the head panel 4 Figure 4 shows a general assembly instruction Be...

Page 9: ...tighten the brackets to the frame until the next step 6 Position the head end brackets over the location labels Ensure that the edge of the bracket is flush with the arrow and is covering the black st...

Page 10: ...ent s hand slipping on the assist device and result in injury or death Warning In the event that a ground wire is disconnected during installation RECONNECT IT NOW Maintenance Inspection Information V...

Page 11: ...from the bed to the side rail Figure 20 Figure 21 Figure 22 A B C D E F G H Figure 19 Maintenance Inspection Information Visually inspect the assist handle and mounting bracket and check for loose har...

Page 12: ...e or defects caused by misuse incorrect handling or installation by a non authorized installer Damage caused by use in unsuitable environmental conditions or failure to maintain the product in accorda...

Reviews: