User Manual

Joerns® Lifting and Repositioning

HPL400 and HML400 Hoyer Patient Lift

To avoid injury, read user’s manual before using.

Joerns Healthcare 1_Art Direction_Joerns Beds_700 Series_UltraCare 8page

5/2007_Page 1 of 8

redefining patient handling

Click here to view detailed product information and pricing

.      

Call 800-397-5899 to order today!

Summary of Contents for HPL400

Page 1: ...ng and Repositioning HPL400 and HML400 Hoyer Patient Lift To avoid injury read user s manual before using redefining patient handling Click here to view detailed product information and pricing Call 800 397 5899 to order today ...

Page 2: ... cradle center may wear prematurely Warn cradle parts could fail resulting in injury to resident caregiver Hanger post should protrude cradle center by 1 16th inch Figure 1 1 Carefully unpack lifter in the following order a Remove the outside top of the box b Remove the three inside top fillers from the box NOTE If the lifter is a power version be careful with the center filler it contains the ele...

Page 3: ...e slings Warning Cradle Center Post must protrude below the bottom of the Cradle Center Pivot Assembly approximately 1 16 Warning If Cradle center Post is flush or recessed excessive wear may have occurred Worn cradle parts could fail resulting in serious injury or death to patient and or caregiver Remove from service and replace worn parts immediately Lock Nut Boom Mast Lock Bolt Shoulder Bolt Ac...

Page 4: ...mergency Lowering Emergency Stop Soft Stop Acoustic battery alarm Troubleshooting Power Unit Symptom Possible Cause Action Power indicator does not light up Not connected to mains The fuse has blown Control box defective Connect to mains Replace fuse if the system is prepared for external fuse replacement or send the system for repair Send control box for repair Power indicator lights up but actua...

Page 5: ...ot charge the battery while someone is in the lifter and especially if the unit is being used in wet or bathing areas Caution By unauthorized opening of the system there is a risk that it may malfunction at a later date Warning Lifters can tip over Keep base widened for stability Never lock brakes or block wheels when lifting someone Brakes are provided for parking To lock casters step on cam leve...

Page 6: ...ation Warning Torn cut frayed or broken slings could fail resulting in serious injury or death to patient Use slings in good condition only Destroy and discard old unusable slings Patient Size Guide A B Weight inch cm inch cm lbs kg Large 28 70 19 47 198 400 90 191 Medium 24 61 19 47 99 210 45 95 Small 18 46 15 38 55 110 25 50 Patient s weight to sling size information is an approximation only Pat...

Page 7: ...and tightness Boom Inspect all bolts and nuts for wear and tightness Inspect to ensure the boom is centered between the base legs Cradle Check hanger post for wear or damage Replace post if wear exceeds 1 16 Check sling attachments for wear or deflection Pump Assembly Tighten seals around piston with spanner wrench provided Pump handle and valve work smoothly Check for oil leakage A light film is ...

Page 8: ...Any alteration modification or repair unless performed by or authorized in writing by Joerns Healthcare will void this warranty This warranty applies in the USA only Check with your supplier to find out if international warranties apply Parts Hoyer Classics Lifts contain various parts that wear from normal use These parts such as rubber parts plastic parts and painted surfaces are not covered unde...