Joerns Hoyer HOY-ADVANCE-E User Instruction Manual & Warranty Download Page 6

294000.10029  Rev. B

6

English

Hoyer

®

4-Point Positioning Cradle

3.  Operating Instructions

1.

  Once the HOY-ADV-6PTWSC spreader bar assembly has been properly installed on the  

  Advance lifter and the sling is in place on the spreader bar:

a)

  Select pounds or kilograms by pressing the “Kg/Lb” switch on the left side of the scale.

b)

  Press the “ZERO” button to set scale system to zero. The digital zero will be displayed for a  

  few seconds, then the LCD will read “0.0”.

2.

  Remove the sling from the spreader bar.

3.

  Position the patient in the sling.

4.

  Reattach the sling to the spreader bar.

5.

  Carefully lift the patient. When the patient’s motion has stabilized press the “On” button.

6.

  Read and record the displayed weight data.  Note: The display will automatically turn off after  

  60 seconds to conserve battery life.  If the display turns off before the weight data has been  

  recorded, simply press the “On” button again.  

7.

  Carefully lower the patient and detach the sling from the spreader bar.  Note: Always reset  

  the scale to “Zero” when changes are made to the sling selection.

4.  Battery Replacement 

The H Electronic Digital Scale is powered by a 9-volt battery. A new battery will provide 

approximately 3000 readings before needing replaced. When a battery needs to be replaced, a 

“B” will appear on the display.

To replace battery:

1.

  Push black pad on back of unit to the left.

2.

  Battery holder will slide out.

3.

  Remove old battery and replace with a new one.

4.

  Push battery holder into scale.

Summary of Contents for Hoyer HOY-ADVANCE-E

Page 1: ...Hoyer Advance Digital Scale User Instruction Manual Warranty To avoid injury read user manual prior to use Hoyer Redefining patient handling...

Page 2: ...Details Contents 1 Safety Warnings 3 2 Assembly Instructions 4 3 Operating Instructions 6 4 Battery Replacement 6 5 Maintenance Calibration 7 6 Warranty 7 Hoyer USA Joerns Healthcare Inc 2100 Design R...

Page 3: ...must use Hoyer slings with Hoyer lifts Using other manufacturer s slings on Hoyer products is unsafe and may result in serious injury to patient and or attendant WARNING Using other manufacturer s pa...

Page 4: ...OLLOW THESE INSTRUCTIONS EXACTLY REMOVAL 1 Support the 6 point spreader bar to prevent the spreader bar from falling unexpectedly The spreader bar can be supported by a chair or table 2 Using the 4mm...

Page 5: ...holes on the spreader bar Note The scale display is to be ori ented away from the lift 5 Insert the quick release pin sleeve through the boom plastic cap and cradle support bracket Fig 2 6 Insert the...

Page 6: ...stabilized press the On button 6 Read and record the displayed weight data Note The display will automatically turn off after 60 seconds to conserve battery life If the display turns off before the we...

Page 7: ...he 6 point spreader bar Tampering with the Hoyerpro Digital Scale and 6 point spreader bar connection may result in serious injury to patient and or caregiver 6 Warranty 1 If allowed by law this warra...

Page 8: ...2012 Joerns Healthcare 294000 10029 Rev B DCO 12 0276 Joerns Healthcare Inc 2100 Design Road Suite 100 Arlington TX 76014 USA Tel 800 826 0270 Fax 800 457 8827 www joerns com...

Reviews: