Method for Connection
Wiring
5.1
5.1.1 Standard type (ARO-FT)
The wire color varies depending on the communication method, UART (3.3V logic) or RS-232C. Make
the connection after checking the communication method.
(RX and TX are from the instrument side.)
Color
UART (3.3 V logic)
RS-232C
Red
Power input (6 - 26 VDC)
Black
GND
Green
FG
Yellow
N.C. (
2
)
RX
White
N.C. (
2
)
TX
Blue
RX (
1
)
N.C. (
2
)
Grey
TX
N.C. (
2
)
5.1.2 Cable type (ARO-FT)
The Cable type is only for RS-232C communication.
(RX and TX are from the instrument side.)
Green
RS-232C
Red
Power input (6 - 26 VDC)
Black
GND
Green
FG
Yellow
RX
White
TX
5.1.3 MCBH type (AROD-FT)
The MCBH type is only for RS-232C communication.
(RX and TX are from the instrument side.)
Pin No.
RS-232C
1
N.C. (
1
)
2
RX
3
TX
4
N.C. (
1
)
5
FG
6
N.C. (
1
)
7
GND
8
Power input (6 - 26 VDC)
Note (
1
): Do not connect wires to N.C. ports. It can damage the instrument.
Note (
1
): 100
kΩ pull
-down
Note (
2
): Do not connect wires to N.C. ports.
It can damage the instrument.