JFE Advantech RINKO FT User Manual Download Page 1

 

 

A-0133-02 

 
 
 
 

Fast optical DO sensor for integration 

 

RINKO FT 

(Model name: 

ARO-FT

AROD-FT

 
 

User’s Manual 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

1.

 

Use after thoroughly reading this manual. 

2.

 

Improper use can lead to accidents. 

3.

 

Safely keep this manual in order not to lose it. 

For safe use 

 

Summary of Contents for RINKO FT

Page 1: ...al DO sensor for integration RINKO FT Model name ARO FT AROD FT User s Manual 1 Use after thoroughly reading this manual 2 Improper use can lead to accidents 3 Safely keep this manual in order not to...

Page 2: ......

Page 3: ...ain unit without changing the float inner pressure for calibration inspection Span zero calibration is possible by the calibration kit optional Table of Contents 1 Terminology and Part Names 2 2 Packa...

Page 4: ...Refers to values such as water temperature and DO calculated from the AD value with calibration coefficients Standard type ARO FT Refers to RINKO FT with standard mounting attachment Cable type ARO F...

Page 5: ...nting attachment Insulation screw Figure 1 2 Cable type ARO FT Figure 1 1 Standard type ARO FT Main Unit Oxygen detection film Temperature sensor Lock nut Locking sleeve LEMO connector Figure 1 3 MCBH...

Page 6: ...r 3m cable with LEMO connector and pigtail end 1 4 User s Manual this manual 1 5 Calibration certificate 1 Note the above list indicates the basic package The content may differ depending on the order...

Page 7: ...1 3 Dummy cap 1 4 User s Manual this manual 1 5 Calibration certificate 1 Note The above list indicates the basic package The content may differ depending on the order 2 Optional parts No Items QTY 1...

Page 8: ...a hazardous situation which if not avoided could result in minor or moderate injury Indicates general precautions Indicates supplements to control method and convenient functions Danger Ensure safety...

Page 9: ...sure to insert the connector completely up to its base It may cause seeping of water into the instrument Do not leave the instrument in a place with high temperature e g under direct sunlight for a lo...

Page 10: ...o the power supply 8 When using the instrument in water it is recommended to use it at 30 C or below When it is used in water at 30 C or above the tracking for correction of temperature change can slo...

Page 11: ...N C 2 Grey TX N C 2 5 1 2 Cable type ARO FT The Cable type is only for RS 232C communication RX and TX are from the instrument side Green RS 232C Red Power input 6 26 VDC Black GND Green FG Yellow RX...

Page 12: ...g hold the metal part and pull it out Pulling only the cable can break it Insert the connector into the main unit until two black O rings are completely hidden Hold it in while going to the next step...

Page 13: ...ut by loosening the locking sleeve Before attaching again apply adequate amount of grease on the O rings When detaching hold the metal part and pull it out Pulling only the cable can break it Insert t...

Page 14: ...ach It can be pulled out by loosening the locking sleeve Before attaching again apply adequate amount of grease on the O rings When detaching hold the connector and pull it out Pulling only the cable...

Page 15: ...by water gently remove it with a soft cloth Rubbing it hard can damage the oxygen detection film 2 After water cleaning remove the moisture either by an air blower or by wiping with a soft cloth 3 Ch...

Page 16: ...ion starts automatically go to 5 If it does not start go to 3 3 Open Computer from the start menu and select the drive with the CD ROM 4 Double click setup exe 5 Screens are displayed in the following...

Page 17: ...dialog may be displayed during installation for security If UAC dialog appears select Yes to continue the installation UAC dialog might not appear depending on the computer setup but this is not abnor...

Page 18: ...is connected to the computer 2 The driver needs to be installed if it displays the screen below Follow the steps below Following procedures are not necessary if the driver is properly installed 3 Ope...

Page 19: ...n below 6 Right click the item with the mark and click Update Driver Software 7 Click Browse my computer for driver software 8 Click Browse button and select the CD ROM or the ARO FT Driver folder ins...

Page 20: ...his driver software anyway 10 If it displays the screen below you have succeeded with the driver installation Check the Device Manager to make sure that the marks such as and have disappeared If the m...

Page 21: ...Manager if it is View by Category Right click Start Computer and search Properties Device Manager b Check Port COM and LPT 3 Start the specialized software FT Calib see 7 4 1 4 Select the port searche...

Page 22: ...7 4 2 Close the software The software can be closed by the following method 1 Close from the File menu The software closes when selecting File Close 2 Close by the button on the title bar The software...

Page 23: ...rer is recommended Place the DO sensor in bubbled water and wait until the value stabilizes When the value stabilizes click the OK button and obtain the value The value can be obtained as many times a...

Page 24: ...lm if the value is not calibrated 7 5 1 How to make air bubbling water for SPAN calibration 7 5 2 How to make sodium sulfite solution for zero calibration Make the sodium sulfite solution referred to...

Page 25: ...rated water This calibration uses air bubbled water as the reference Enter Atmospheric Pressure at the time of calibration RMDO Reference Material DO Under development A reference material can be used...

Page 26: ...ing the Refresh button Click the Select button to select the file to be updated After clicking the Start button two confirmation screens will appear Click the OK button to start the update Restart the...

Page 27: ...ion 1 Start the software FT Calib see 7 4 1 2 Select Help Version information The Version information window will be displayed The content and specifications may change for correction and improvement...

Page 28: ...lose without saving Item Description Log Folder The folder where the logs of calibration and communication are saved Timeout Time length of response to communication commands If time out occurs freque...

Page 29: ...gs are displayed 7 7 2 Display of precautions for calibration A caution for calibration is displayed when starting the software The display of the caution can be changed by Option see 7 7 The display...

Page 30: ...he guidance 7 8 2 Uninstall using the Programs and Features From Start menu click Control Panel From the Control Panel window select Programs and Features Select FT Calib from the list by clicking it...

Page 31: ...rong noise source is not in contact with the main unit or the lead wire Check if the measurement value is obtained after the preheat time 5 seconds Check if there is any large scratch on the oxygen de...

Page 32: ...Set up the connection port see 7 3 The connection port is wrong If the computer has multiple ports a port which is not connected to the instrument may be set up Confirm the port number see 7 3 Time ou...

Page 33: ...interval 1 s Pre heat time 5 s Model name ARO FT AROD FT Depth rating 2 000 m depth equivalent 6 700 m depth equivalent Material Housing Titanium grade 2 Insulation screw POM Housing Titanium Ti 6Al...

Page 34: ...8 1 GB or larger 32 bit 2 GB or larger 64 bit HDD Hard disk with space more than 1 GB About 30 MB is needed for installation Resolution 1024 768 pixels or greater Port USB 1 Other NET Framework 3 5 S...

Page 35: ...33 Drawing 9 4 Dimensions are in mm Figure 9 1 Standard type ARO FT Figure 9 2 Mounting part of standard type ARO FT Insulation screw and O rings Sensor side Mounting hole Platform side...

Page 36: ...34 Dimensions are in mm Figure 9 3 Cable type ARO FT Figure 9 4 MCBH type AROD FT...

Page 37: ...compensated oxygen concentration pc is calculated from the equation below pc 1 Pressure MPa Calibration coefficient The salinity compensated oxygen concentration sc is calculated by multiplying the f...

Page 38: ...damage due to external causes such as fire earthquake flood lightening other natural disasters public nuisance abnormal voltage corrosive gas organic solvent and chemical solution f Malfunction and da...

Page 39: ......

Page 40: ...p mode The checksum is the least significant byte of ones complement of the sum of the values of Request or Response and comma The checksum should be converted to 2 byte ASCII For example In case of a...

Page 41: ...FF when receiving the command it takes 5 seconds for the DO value to stabilize For that reason DO NOT USE this command during the sleep mode because the RINKO FT again enters sleep mode immediately af...

Page 42: ...e output value 5 means 50 ms In order to calculate the LED accumulated time in seconds LED accumulated time s decimal expression for LED accumulated time HEX 100 F data format F is a real number F is...

Page 43: ...ternally calculate the AD values of DO The physical values of DO can be converted from the AD values of DO using the equation provided in the manual section 9 5 First response from the sleep mode RINK...

Page 44: ...d o n HEX Temp HEX DO Temp 4 unfixed T AD value DO 4 DO AD value 15 16 17 18 19 20 21 22 23 24 25 26 27 28 LED 8 LED accumulated time HEX LED CR LF 0 1 2 3 4 5 6 7 8 9 11 tdona Request t d o n a CR LF...

Page 45: ...e B Temp 4 unfixed T AD value DO 4 DO AD value 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 Phase B 4 Blue phase AD value HEX Phase R HEX Amplitude B HEX Amplitude R Phase R 4 Red phase AD value Ampli...

Page 46: ...6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 d3 F d 3 F d3 coefficient CR LF d3 Max unfixed DO d3 coefficient coefficient 16 Uchida et al 2013 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 1...

Page 47: ...Calibration 10 unfixed Date of T calib date 0 1 2 3 4 5 6 7 23 wu Request w u CR LF 0xE7 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 24 wu State Response w u n o r m a l CR LF State normal 0x32 0 1 2 3 4 5 6 7...

Page 48: ...responds after error 0003 see Chapters 3 and 8 in appendix 1 Example 1 Sample T and DO at 10 s or more of sampling interval It is much better to power on and off RINKO FT for each measurement to save...

Page 49: ...y host controller Command qs No response Power OFF RINKO FT End OK Repeat sampling Yes No Example 2 Sample T and DO at 5 s or less of sampling interval In this case RINKO FT is necessary to be continu...

Page 50: ...Manufacturer 3 48 Takahata cho Nishinomiya Hyogo 663 8202 Japan Phone 81 798 66 1783 Fax 81 798 66 1654 URL http www jfe advantech co jp E mail ocean jfe advantech co jp...