JSM6010A SEM Short Manual  

(modified on Feb 10, 2016) 

 
Start SEM instrument 

 

1.

 

Turn on the key-switch to START and hold for 10 seconds 

2.

 

Turn on the personal computer and  Login PC with user name: NTNU and password: 
SEMUser 

3.

 

Double  click  the  shortcut  icon  “In  TounchScope  (Navi)”,  which  is  full  “dialogue 
version” with step-by-step guides; the more professional version with shortcut as “In 
TounchScope” (NB, see Step 4 below) 
(Laptop screen ratio is not compatible with SEM program ratio. Hold on the Mouse 
button to move the whole program up and down) 

4.

 

For using “In TounchScope”, if the sample top is higher than the holder top level, it 
has to move down Specimen Height Handle correspondingly to protect BSE detector!!!  

5.

 

Click “Set specimen”. After about 70s, the specimen chamber is vented to atmospheric 
pressure. The observation preparation window is displayed.  

 

Sample Loading and Exchange for using 

In TounchScope (Navi)

 

 

1.

 

Load a specimen according to the steps displayed on the window:  

a.

 

Draw out the specimen stage to its end-stop position to load the specimens. 

b.

 

Input  the  protruding  specimen  height  into  the  dialog  box  if  the  specimen 
protrudes from the upper surface of the specimen holder. 

c.

 

Selecting the specimen holder type. Click “Select” to select and click “Set”  

d.

 

Adjust the stage Z-axis to 10 mm. Check the stage positions are R=0

o

 and T=0

o

 

e.

 

Close the specimen chamber 

2.

 

Click the “Observation Conditions” button and a few recipes are appeared. Click the 
appropriate button suiting for your sample. 

3.

 

Setting the observation conditions. 

4.

 

Click  the  “Start”  button  and  the  chamber  starts  evacuating  and  image  automatically 
appears when the evacuation is completed. If the gun filament is not properly adjusted, 
the image may not appear. For details how to adjust the filament, refer to page 4-82 in 
the INSTRUCTIONS. 

 

To exchange specimens immediately: 
1.

 

First click HT icon “on” to turn off the HT, and then click “VENT” button and click 
the “Yes” button in the dialog box.  

2.

 

Draw out the specimen stage to its end-stop position to load the specimens. 

3.

 

Input the protruding specimen height. Then, set the Z axis to the value in the message 
and click the “Close” button. 

4.

 

Closing the specimen chamber, click the “EVAC” button on the icon window 1. 

5.

 

Click the “YES” button in the dialog box. 

6.

 

Click HT icon “off” to turn on the HT. 

 

Abbrev.  

AFS, Auto Filament Saturation 

 

AF,    Auto Focus 

 

 

AGC, Auto Gun Control 

 

 

ACB, Auto Contrast&Brightness 

 

 

AGA, Auto Gun Alignment  

 

AS, Auto Astigmatism correction 

 

Using  IPad, 

Log-in  Apple  IPad  with  PIN  6010.  Go  to  IPad  Wi-Fi  Setting,  to  click  Router 

ZyXELDD6014  link;  then  back  to  Home  Screen,  to  click  SEMController  icon  to  start  SEM 
Controls, where to control AF, ACB and AS, together with corresponding manual controls.

 