1
Experimental Conditions
For the preparation/coating of conodont specimens, the coating conditions are as follows:
Voltage: 1200 V (5–10 mA) and coating time of 4 min
Process is stopped and stubs are rotated by ~90°
Continue coating for 4 min
In case of severe charging of conodont specimens, the following coating conditions have been followed:
Coating time set to 10 min
Process is stopped and stubs are rotated by ~120°
10 min coating time
Process is stopped and stubs are rotated by 120° more (total: 240°)
10 min coating time
Coating Time: 10 min x 3 times
Summary of Contents for JFC-1100