JEOL 1010 Quick Start Manual Download Page 5

Version 2017-01-19 for GI use 
 

 

In case the alignment of the microscope is lost, which will for example become 
evident because the beam has an oval shape (astigmatism), the original settings can 
be recalled. Ask assistance

. [Note for assistants:  to the hardware keyboard (the one 

going along with the small monitor), press ESC to make sure to start with a clean line 
and type: ufc o 1 (which is ufc in small letters,  space, letter o , not zero, in small font, 
space and then 1). Press Enter. ]

 

 

To acquire an image, press on the right camera icon of intelligent exposure. The 
entire gray levels displayed on the histogram are automatically recorded. However, 
the brightness settings should be adjusted in such a way that the histogram is 
centered and as broad as possible so that many gray levels are recorded, but also 
the exposure time should be as short as possible, in order to prevent vibrations to 
bias the image quality. Note that there is a slight difference in eveness ofillumination 
for direct vision under the column or through the camera (use brightness to 
compensate). 

 

In case automatic balancing of the 
camera is not suitable, for example 
because the interesting parts in the 
specimen are very dark or very bright, 
one can apply (automatic, fixed scale or 
manual) exposure on a Region of 
Interest around these part (click the 
green screen icon within the Intelligent 
exposure pop-up window and drag a 
ROI). Manual exposure, manual 
minimum and maximum setting and 
manual gamma adjustment is also possible. Ask Geert-Jan for the various options. 
 

 

For saving images, enter the correct magnification (mag x can be read on the small 
hardware monitor) and image name in the own folder (D:>Users> path). Scale bars 
are automatically burned into the image. (Cancel the second pop-up screen). 

Images 

can be uploaded to a server at the end of a session.

 

 

To finish a session, go back to small magnification (lower diaphragm handle to the 
right). Finish after having taken one picture to allow the charge on the chip of the 
camera to be withdrawn and switch of the filament 

2

 

 

To unload the specimen holder, always check first (once again) that the 

FILAMENT 

IS  OUT (!)

Gently pull the rod to the final stop. Turn one quarter anti-clockwise. 

Release the rod by holding one finger against the goniometer to counter the vacuum 
sucking in a controlled manner and pull the rod out.

 If you are the last user of the day, 

switch also the HT 

1

 out. Exit the program. 

 

Place the grid rod holder on the standard and release the grid cover with the wing 
key. Allow the grid to fall on a clean piece of paper and lift it with tweezers to set it 
back in the grid box. Close the cover and secure the wing key in its cavity.  

 

H

 

ALIGNMENT JEOL 1010 TEM (GI) 

HOW TO CHECK THE SATURATION OF THE TEM FILAMENT? 

1. 

Start the High Tension (HT) and the filament (Fil) 

Summary of Contents for 1010

Page 1: ...he slit over the cover grab the cover with the red dots oriented in the long axis turn a quarter and gently open it Insert the grid specimen up Close the cover and remove the wing key Check that all c...

Page 2: ...n the goniometer Gently press the grid rod straight into the opening To make contact press the rod cylinder metal color in ours with a flat hand palm a green light lights line Wait until the green lig...

Page 3: ...Version 2017 01 19 for GI use 2 Viewing specimen acquiring images right panel left panel and central base of the column with xy navigation...

Page 4: ...tor use switch 4 MAG 2 is for pre set magnifications If possible use standard magnification steps to better compare images XY navigation with 14 and 15 Spread the bundle as evenly as possible just ove...

Page 5: ...these part click the green screen icon within the Intelligent exposure pop up window and drag a ROI Manual exposure manual minimum and maximum setting and manual gamma adjustment is also possible Ask...

Page 6: ...lower condenser diaphragms by switching the level c to the right 3 Go up and down with brightness 12 Adjust over and underfocus If the A well centered filament gives a centered beam position swings ad...

Page 7: ...he beam with brightness 7 Reinsert the upper diaphragm c and the lower diaphragm HOW TO ADJUST THE UPPER CONDENSER DIAPHRAGM 1 Switch HT and Fil on Go to LOW MAG and spread the beam 2 The upper diaphr...

Page 8: ...s 5 Use screw b first completely in until stop Be careful not to to prevent the diaphragm to fall into the column 6 FRESNEL FRINGES ADJUSTMENT 1 Insert a carbon test grid ASK Geert Jan Put HT and Fil...

Page 9: ...g in the TEM a beam alignment should be performed prior to use A tool for this alignment is a holey grid A holey grid is a TEM grid support coated with a thin plastic film and a stabilizing carbon lay...

Page 10: ...lie stated that the wavelength of an electron beam is a function of the accelerating voltage used By increasing the accelerating voltage a shorter wavelength is obtained The shorter wavelength is appl...

Page 11: ...increase accelerating voltage improve the vacuum and or use a thinner specimen An astigmation occurs when a lens field is not symmetrical in strength but weaker in one plane than another Astigmation...

Page 12: ...nd right to center the spot right on the small dot on the phosphor plate Spread the bundle with the brightness knob Center the position of the spread bundle using the lateral and front knobs of the up...

Page 13: ...tween Brightness and Def Apply iterations until the rim is central respect to the hole When ready OBJ STIG off Focus And check with the wobbler that not shift occurs BUGS TROUBLE SHOOTING When the swi...

Page 14: ...holder Each position can be found by switching SPEC SEL left bottom of the column However as a routine only the position 1 is used When a very bright spot that can not be moved is present in the mid o...

Page 15: ...Version 2017 01 19 for GI use HT and FIL on again Bugs...

Reviews: