Operation
31
Calibration range
After selecting this option, the following window appears:
Select the measuring range by moving the cursor to the desired option
using the arrow keys. Subsequently, press
.
Depending on the VARIOSCAN
high resolution
model, the following
measuring ranges are available:
Aperture
B1
B2
B4
B8
B16
Fi
3021 (°C) -40...100 40...400 100...800 200...1200
3022 (°C) -10...100 40...200 100...400 300... 800 500...1200 -10...100
On Model 3022, selection of „Fi“ moves a bandpass filter into the
optical path. This filter has a transmission range of 3 ... 5 µm. Use this
setting when you work with this camera outdoors in the sunshine.
Emissivity
Entry of the global emissivity value. This value is taken into account in
temperature calculation. The parameter is material-specific considering
the emissivity of the object.
Environmental temperature
This is the temperature of objects in the ambience of the measured
object emitting heat radiation that might be reflected at the surface of
the measured object. This radiation will affect the object temperature
measured by VARIOSCAN
high resolution
only if the set emissivity is
less than 1.
InfraTec
Summary of Contents for VARIOSCAN high resolution 3021
Page 1: ...VARIOSCAN high resolution Models 3021 3021 ST and 3022 Manual JENOPTIK Group I n f r a T e c ...
Page 10: ...Technical Description 10 I n f r a T e c ...
Page 12: ...Technical Description 12 I n f r a T e c ...
Page 14: ...Technical Description 14 Total weight of system in carrying case approx 11 kg I n f r a T e c ...