Serial interface and communication software
LDM52 OEM, Release 003, Revision 000
45
7
Serial interface and communication software
7.1
Transmission protocol
Interface settings: Asynchronous, 8 data bits, no parity, 1 stop bit
Transmission protocol format / syntax: 7 bit ASCII
Proprietary transmission protocol
Commands are case-insensitive (NO differentiation between lower and upper case).
Decimal separator in the output of figures is the dot “.” (0x2E).
The terminator of a command (sending command) is the enter key (0x0D, 0x0A) or Car-
riage Return (0x0D) or Line Feed (0x0A)
Where parameters have several values, they are separated by a space (0x20).
The response to commands with parameters is the respective command including the pa-
rameters.
The response to commands without parameters is the respective command including the
current parameters.
The response to commands with parameters outside of the valid value range is the re-
spective command including the current parameters.
The response to unknown commands and faulty parameter formats is a “?” (ox3F).