JENCO VisionPlus pH 618 Operation Manual Download Page 10

9

ERROR DISPLAYS AND TROUBLESHOOTING 

pH LCD 
Display 

ATC
Display 

DISPLAY
 Mode 

Possible cause(s) 
[Action(s)]

Any
Value 

"OVER"

Measure

a. Temperature over 0 to 

99.9°C range. 

[Bring solution to a 
lower/higher
temperature.]
[Replace temperature 
probe.]
b. No temperature 

sensor.

[Use a temperature 
probe.]

"OVER" 0.0~ 

99.9°C  Measure

pH value over 0.00 
to14.00pH range . 
[Over Range or 
Recalibrate.]

"OVER" 

0.0 ~ 60.0°C

pH CAL 

a.pH-

STAND

b.pH-

SLOPE

a. Offset @ 7.00pH: 

mV>90mV

    [email protected]

mV>98.3mV

b. New slope>ideal 
slope by 30%
[Use a new buffer 
solution.]
[Replace electrode.] 

"OVER" 

0.0 ~ 60.0°C

pH CAL 

a.pH-

STAND

b.pH-

SLOPE

a. Offset @ 7.00pH: 

mV<-90mV

Offset @ 6.86pH: 
mV< -81.7 mV 

b. New Slope<ideal 

slope by 30%

[Use a new buffer 
solution.]
[Replace electrode.] 

Any
Value 

"OVER"

pH CAL 

a.pH-

STAND

b.pH-

SLOPE

Buffer temperature over 
0 to 60 °C range. 
[Bring buffer 
temperature within 0 to 
60°C range .] 

Summary of Contents for VisionPlus pH 618

Page 1: ...pH618 Operation Manual MODEL pH 618 Microcomputer Based pH Temperature Pocket Meter...

Page 2: ...3 B VisionPlus Graphical Display 3 OPERATION MODES AND KEYPAD OPERATIONS 4 A Operation Modes 4 B Keypad Operations 4 USING VISIONPLUS pH 618 5 A Power ON OFF 5 B Set Date and Time 5 C Select Buffer 6...

Page 3: ...ure and gently take off the cover A small screwdriver is provided to easily remove the screws 2 Remove all of the old batteries and insert a new set of batteries ensuring the polarities are correct B...

Page 4: ...Electrode collar 5 Electrode ATC assembly Electrode cap is not shown B VisionPlus Graphical Display 1 DATE MM DD YY 2 HOLD mode indicator 3 MEASURE mode indicator 4 CALIBRATION buffer s 5 TIME HH MM...

Page 5: ...e Set Mode is used to change date and time 6 Buffer Select Mode Buffer Select Mode is used to select the buffer set which can either be 7 00 7 00 4 01 10 01 or 6 86 6 86 4 00 9 18 B Keypad Operations...

Page 6: ...nd Time Set 0 second Decreases current number SCROLL Buffer Select 0 second Selects new buffer set USING VISIONPLUS pH 618 A Power ON OFF Press HOLD key for 2 seconds or more The unit will turn on and...

Page 7: ...pH meter will perform end point sensing to determine when the calibration reading is stable Meter is calibrated when WAIT icon disappears and the unit will automatically enter the next point calibrat...

Page 8: ...d Mode or the Measure Mode press STORE key for 2 3 seconds The DATA SAVE and M XX icons will appear indicating the reading has been saved and stored in memory location XX The unit will automatically r...

Page 9: ...collar 3 Insert a new electrode make sure the electrode fit back into the meter correctly 4 Screw back the electrode collar B Replace Calendar Battery 1 Unscrew the electrode ATC assembly as shown in...

Page 10: ...er 0 00 to14 00pH range Over Range or Recalibrate OVER 0 0 60 0 C pH CAL a pH STAND b pH SLOPE a Offset 7 00pH mV 90mV Offset 6 86pH mV 98 3mV b New slope ideal slope by 30 Use a new buffer solution R...

Page 11: ...ode Slope recognition 30 at pH 4 00 4 01 9 18 and 10 01 Input impedance 1012 Temperature Temperature sensor Thermistor 10 k at 25 C Temperature unit Default in factory General Power Meter LR44 x 4 Tim...

Page 12: ...ng the reason for return For your protection items must be carefully packed to prevent damage in shipment and insured against possible damage or loss Jenco will not be responsible for damage resulting...

Reviews: