JENCO VisionPlus EC 330 Operation Manual Download Page 5

4

BEFORE YOUR FIRST USE

A. Insert Batteries

1. Remove the battery cover

at the top of the unit.

2. Insert the set of batteries

(included) ensuring correct

polarities.

3. Securely replace battery

cover.

B. Soak the Electrode

1. Remove the electrode cap covering the VisionPlus EC 330

meter.

2. Soak the electrode in distilled water for 10 minutes before

first use or after storage.

C. Preparing Standard Solutions

Suitable conductivity standards are available commercially or the

user can prepare them using research grade reagents.
Here are some standard solutions the user can prepare to

calibrate the probe of the model EC330.

1. Standard solution of 1413uS at 25

: Accurately weight out

0.746 grams of research grade dried Potassium Chloride

(KCL). Dissolve in 1000ml of distilled water.

2. Standard solution of 12.90mS at 25

: Accurately weight

out 7.4365 grams of research grade dried Potassium

Chloride (KCL). Dissolve in 1000ml of distilled water.

[

Note:

You can store the remaining solution in a plastic container

for one week but the air space between the cap and the solution

must be kept to an absolute minimum. Storing the excess solution

below 4

can increase the storage life. If you have any doubt of

the accuracy of the stored solution, a fresh batch should be

prepared.]

1

2

Summary of Contents for VisionPlus EC 330

Page 1: ...Operation Manual MODEL EC 330 Microcomputer Based Conductivity TDS Temperature Pocket Meter EC330...

Page 2: ...ypad Operations 3 BEFORE YOUR FIRST USE 4 A Insert Batteries 4 B Soak the Electrode 4 C Preparing Standard Solutions 4 D Setup and Calibrate the Electrode and Meter 5 USING VISIONPLUS EC 330 5 A Power...

Page 3: ...materials should be saved until satisfactory operation is confirmed VISIONPLUS EC 330 OVERVIEW A Meter Description 1 Battery cap 2 LCD screen 3 Keypad 4 Electrode collar 5 Electrode ATC assembly Elec...

Page 4: ...alibration Mode Calibration Mode is used to perform 1 or 2 point calibration 3 Hold Mode Hold Mode is used to display the locked reading for increased ease of use B Keypad Operations Key Operation Mod...

Page 5: ...ome standard solutions the user can prepare to calibrate the probe of the model EC330 1 Standard solution of 1413uS at 25 Accurately weight out 0 746 grams of research grade dried Potassium Chloride K...

Page 6: ...ater and immerse them in the standard solution of 1413uS The temperature displayed is the solution temperature 2 Press Cal key to initiate calibration The CAL icon will appear when the main display sh...

Page 7: ...the reading 2 Press Hold key again to unlock reading and the unit will return to Measure Mode The unit is now ready for another measurement REPLACE ELECTRODE 1 Unscrew the electrode collar to remove...

Page 8: ...LCD screen The instrument can operate within specifications for approximately 2 3 hours after low battery indicator appears 1 Take off the battery cover 2 Remove all of the old batteries and insert a...

Page 9: ...higher temperature Replace electrode ATC assembly OVER 60 0 99 5 C Measure Temperature 60 0 C over the temperature compensation range Bring solution to a lower temperature OVER 0 0 60 0 C Measure The...

Page 10: ...0 10 g L 7 50 10 00 g L Temperature Range Resolution Accuracy 0 0 to 99 5 C 32 to 212 F 0 5 C 1 F 0 5 C 1 F Conductivity Temperature compensation Temperature Coefficient Reference temperature AUTO 0...

Page 11: ...product problem is due to deviations or customer misuse Out of warranty products will be repaired on a charge basis RETURN OF ITEMS Authorization must be obtained from one of our representatives befor...

Page 12: ...om Jenco Electronics Ltd 6F NO 81 Sec 2 Chang an E Rd Jhongshan District Taipei City 104 Taiwan TEL 886 2 2508 2928 FAX 886 2 2508 2938 E Mail sales JENCO com tw Website www JENCO com tw Shanghai Jenc...

Reviews: