JENCO pH610N Operation Manual Download Page 2

[ Bring solution to a lower 
temperature ]

[ Bring solution to a higher 
temperature ]

[ pH610N

Recalibrate with 

standard buffer solutions and take 
the measurement again ]

[ ORP650N

Measure Standard 

ORP solutions. If the readings are 
out of range, contact vendor ]

[ Calibrate with a new set of buffer 
solution or replace the tester ]

[ Calibrate with a new set of buffer 
solution or replace the tester ]

a. Offset calibration reading for 

(

 

7.00 pH

 

/

 

6.86 pH

 

) is greater than 

(

 

100 mV

 

/

 

108.3 mV

 

)

b. New slope > ideal slope by 30

 

%

a. Offset calibration reading for 

(

 

7.00 pH

 

/

 

6.86 pH

 

) is lower than 

(

 

-100 mV

 

/

 -

91.7 mV

 

)

b. New slope < ideal

slope by 30

 

%

Warranty

Jenco warrants this product to be free from significant deviations in material and workmanship for a period of 1 year from date of purchase. If repair or 
adjustment is necessary and has not been the result of abuse or misuse, within the year period, please return-freight-prepaid and the correction of the 
defect will be made free of charge. If you purchased the item from our Jenco distributors and it is under warranty, please contact them to notify us of 
the situation. Jenco Service Department alone will determine if the product problem is due to deviations or customer misuse. 

Out-of-warranty products will be repaired on a charge basis.

About Jenco

Jenco is a family-operated manufacturer with 45 years of specialization in designing and producing water quality instruments. For more information 
about the Wand tester or other Jenco products, please visit 

https://jencoi.com

7

Calibrate pH (for pH610N)

pH610N defaults to USA buffer set ( 4.01, 7.00, 10.01 ).  To 
change the buffer set, while the unit is off press and hold the 
“CAL/HOLD” key while pressing the “On/Off” key to turn on the 
tester.

 

Note

Only repeat this procedure to change the buffer set.

1. Remove the electrode cap of the instrument and rinse the 

electrode with distilled water.  Place the electrode into the 
6.86/7.00 pH buffer.  

2. Press and hold the “Cal” button for 2 seconds to enter the 

calibration mode.

3. “Cal” is displayed on the LCD to indicate the current mode, 

and either 6.86 or 7.00 pH buffer value is displayed.

4. Blinking “HOLD” is displayed to indicate the tester is 

stabilizing.  When the measurement is stabilized, the tester 
automatically stores the calibration value and the first 
calibration point is complete.

5. “udr” is displayed while the tester waits for the next 

calibration point, which defaults to pH buffer 9.18/10.01 or 
pH buffer 4.00/4.01.

6. Press “CAL/HOLD” key to exit calibration and save the first 

calibration data point, or rinse the electrode with distilled 
water and place it into either pH buffer 4.00/4.01 pH or  
9.18/10.01 to continue to calibrate the second calibration 
point. 

7. Blinking “HOLD” is displayed to indicate the tester is 

stabilizing.  When the measurement is stabilized, the tester 
automatically stores the calibrated value and the second 
calibration point is complete.

8. "Cal" and "HOLD" is no longer displayed to indicate the 

second point calibration is completed, and the top part of 
the LCD displays the two calibrated buffer values.

Note

pH 610N must be calibrated before first use. 

               ORP650N does not require any calibration.

9

Hold Data

1. When the pH/ORP reading is stable, press “Hold” key to hold 

and lock the display to the latest measurement values.

2. Press “Hold” key again to release and unlock the display and 

the tester returns to “Measure Mode”. The unit is now ready 
for another measurement.

Specifications

8

Measure

1. From “Measure Mode”, dip 

the electrode end of the 
tester into the test solution.

Error Displays and Troubleshooting

pH Buffer Recognition

pH Electrode Offset Recognition Range

pH Electrode Slope Recognition Range

pH 7.00, 4.01, 10.01 or pH 6.86, 4.00, 9.18

pH 4.00, 4.01, 9.18, 10.01

±30 %

pH 7.00

±100 mV

pH 6.86

-91.7 mV to 108.3 mV

 

pH610N

 

ORP650N

 

Parameters

 

pH / Temp

 

ORP / Temp

 

Range

 

0.00~14.00 pH

 

±1000 mV

 

Resolution

 

0.01 pH

 

1 mV

 

Accuracy

 

±0.05 pH

 

±3 mV

 

Temp.  Range

 

0.0~50.0 °C (32 .0~122.0 °F)

 

Temp.  Resolution

 

Temp.  Accuracy

 

±0.5 °C / ±1 °F

 

Input

  Impedance

 

3X10

11

Ω

 

 

Power 

LR44 x 4 

Battery 

 

>150 hours

 

Ambient 
Temperature 
Range

 

 

Waterproof

 

IP67

 

Dimensions

 

163 x 43 x 20 mm

 

Weight

 

65

 

g

 

 

Life

0.1 °C ( ≤

 

99.9 °F, 0.2 °F ; >

 

99.9 °F, 1 °F )

0.0~50.0 °C