instruction sheet in case you need to reload these calibration
coefficients. New probe/cable assemblies have the sensor cap
installed and coefficients loaded into the probe at the factory. When
a cable is installed, the coefficients are automatically sent to the
instrument.
Preparing the probe for the first time:
1.
Remove the probe guard from the probe by turning it
counterclockwise.
2.
Remove the red storage cap which contains a moist
sponge from the end of the probe by pulling it straight off
the sensor. Save this to use later for long term storage.
3.
Reinstall the probe guard by sliding it carefully over the
sensor and then threading it onto the cable/probe assembly
with a clockwise rotation (Figure 5).
4.
Locate the gray calibration/storage sleeve that was
shipped with your probe/cable assembly. Moisten the
sponge in the gray calibration/storage with a small amount
of clean water.
5.
Slide the calibration/storage sleeve over the probe guard to
keep the probe in a moist environment for storage and
calibration (Figure 6). It is important to always keep your
sensor in a moist environment so the sensor cap does not
dry out.
6.
Connect the cable to the instrument by aligning the
connectors, pushing them together and then screwing
down the retainer clockwise.