17
8.
Temp. control settings page
a. Screen illustration
TEMP.CONTROL SETTINGS
1.RELAY 5 :HIGH
2.SET POINT :100.0 C
3.HYSTERESIS:EDGE
* SAVING *
4.HYSTERESIS: 1.0C
1)
RELAY 5
– The temperature has only one relay to control. The user needs to set what
control type it will use,
HIGH
or
LOW
type. (In HIGH-type,
RELAY 5
will turn
ON
if the
temperature is greater than the
RELAY 5
setting. In LOW-type, the
RELAY 5
will turn
ON
if the temperature is less than the
RELAY 5
setting which is modified by the
hysteresis value and hysteresis mode.
2)
SET POINT
- This is the
RELAY 5
control set point.
[
Note
: The user can set the
SET POINT
from 0.0 to 199.9 °C, but be reminded that
the range of the temperature is –10.0 to 120 °C.]
3)
HYSTERESIS
- This is the hysteresis mode for temperature relay. The user can
choose between “
CENTER
” or “
EDGE
”.
4)
HYSTERESIS
-This is the actual value of the hysteresis. The user can change this
value from 0.1 to 19.9 °C.