![JENCO 3173 Operation Manual Download Page 7](http://html.mh-extra.com/html/jenco/3173/3173_operation-manual_641896007.webp)
6
MODES OF
THE
METER
This instrument has four modes as follow:
Temperature
is always displayed in all modes.
1.
Conductivity
(uS or mS):
A measurement of the conductive material
in the solution with no regard to
temperature.
2.
Compensated Conductivity
(uS or mS):
Also known as temperature compensated
conductivity which automatically adjusts
to a calculated value which would have
been read if the sample had been at 25
℃
( or some other reference temperature
which the user chooses). See section
CALIBRATION SETUP
.
3.
Salinity
(ppt):
A measurement of the salinity in the
solution, the calculation is based on the
Conductivity and Temperature.
4.
TDS
(ppm or ppt):
A measurement of the total dissolved
solids (TDS), the calculation is based on
the Compensated Conductivity.