Chapter 3
DS3 Test Commands
Description of DS3 test commands
CT-650 Command-Line Reference Guide
Release 9.4
43
Example
The following command monitors a dual-FAD 3/3 DCS and perform an
EF split.
DS3_Split EF_Split Idle_Code
FEAC_Alarm
Selects the FEAC alarm to be transmitted in C-bit or multiplexed C-bit
framing mode. The alarm is only transmitted for the required duration
of two seconds. FEAC alarms can only be transmitted when the RTU
idle condition is set to
IDLE
.
Syntax
FEAC_Alarm <
> <
>
Example
The following command transmits a DS3 FEAC alarm indicating
“service-affecting equipment failure” from the connected side.
feac_alarm connected ds3_sa_eq_fl
Remark
For a detailed listing of messages and alarm codes, see
.
Monitor
Starts a new test.
Syntax
Monitor <
> <
<
Remarks
– The parameter
fad_type
is available only for DCSs that are
configured for single and dual FAD test port types.
– If the
framing_format
is
AUTO
, the CT-650 automatically
synchronizes to the received signal framing. If one of the other
framing_formats
is chosen, the transmitter is set up based on
the RTU idle condition default (Idle or Thru) and the specified
framing is saved to be implemented when the
DS3_Split
command is issued.
Summary of Contents for CT-650
Page 1: ...CT 650 Wideband Test Unit Release 9 4 Command Line Reference Guide ...
Page 2: ......
Page 3: ...CT 650 Wideband Test Unit Release 9 4 Command Line Reference Guide ...
Page 4: ......
Page 22: ...Contents xxii CT 650 Command Line Reference Guide Release 9 4 ...
Page 240: ...Appendix D Customer Services and Support Training options 214 CT 650 User s Guide Release 9 3 ...
Page 247: ......