![JDS Uniphase CellAdvisor JD780A Series User Manual Download Page 558](http://html1.mh-extra.com/html/jds-uniphase/celladvisor-jd780a-series/celladvisor-jd780a-series_user-manual_3602616558.webp)
JD780A Series Analyzers
558
Using Power Meter
Procedure
1.
Connect a MP-series optical power sensor to the
USB Host
port of the JD780A series as shown
in Figure 295.
The OPS indicator appears in the system status bar.
2.
Connect the optical cable from the output port of BTS to the power meter.
Figure 295
Optical power sensor connection
JD785A/JD786A
JD788A
Selecting measurement mode
Procedure
1.
Press the
MODE
hard key.
2.
Select
More (1/2) > Power Meter > Optical Power Meter
.
The Optical Power Meter screen appears.
Setting measurement parameters
Procedure
1.
Press the
MEASURE SETUP
hot key.
2.
To select your desired wavelength for testing, do one of the following:
–
To let your instrument detect the wavelength automatically, toggle the
Wavelength
soft key
and select
Auto
.
–
To select a wavelength from the wavelength list, complete the following steps:
a.
Toggle the
Wavelength
soft key and select
Manual
.
b.
Press the
Stdwave
soft key.
The wavelength list box appears.
c.
Select the wavelength you want from the list, and then tap the
Apply
button.
3.
Optional
. To edit the wavelength list, complete the following steps:
a.
Press the
Stored Wavelength
soft key.
The Stored Wavelength dialog box appears.
b.
Do the following:
–
To add a new wavelength, enter a value between 780 and 1650 in the input box,
and then tap the
Add
button.
–
To edit an existing wavelength, highlight a wavelength in the list, enter a new value
between 780 and 1650 in the input box, and then tap the
Edit
button.
–
To delete an existing wavelength from the list, highlight a wavelength in the list, and
then tap the
Delete
button.
c.
Tap the
Apply
button to save your changes.
Summary of Contents for CellAdvisor JD780A Series
Page 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Page 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Page 525: ......
Page 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...