![JDS Uniphase CellAdvisor JD780A Series User Manual Download Page 495](http://html1.mh-extra.com/html/jds-uniphase/celladvisor-jd780a-series/celladvisor-jd780a-series_user-manual_3602616495.webp)
JD780A Series Analyzers
Using Cable and Antenna Analyzer
495
Calibration accessories (optional).
Mechanical Y-Cal kit or electronic EZ-Cal kit
Phase-stabilized test cable
1-port calibration
Measurement modes that need the 1-port calibration are Reflection, DTF, Cable Loss, 1 Port Phase,
and Smith Chart.
It is recommended that you perform an O-S-L calibration right at the Reflection/RF Out port of the
instrument without using an extension cable in order to minimize a measurement error. If using an
extension cable is inevitable, you need to use a phase stable cable and perform the O-S-L calibration
at the open end of the extension cable.
NOTE
If the O-S-L calibration is done at the end of the port extension cable for DTF
measurement, the length of the port extension cable is compensated automatically and is
not included in the distance to the point of discontinuity.
Bending or moving the extension cable while making a measurement may cause errors in
the measurement.
O-S-L calibration using mechanical Y-Cal kit
illustrates the connection when a port extension cable is used for calibration.
Figure 246
1-port calibration connection with mechanical Y-Cal kit
Procedure
1.
Press the
MEASURE SETUP
hot key.
2.
Press the
Calibrate
soft key.
The on-screen instruction for the calibration appears.
3.
Connect the
OPEN
connector of the CAL Kit directly to the
Cable and Antenna Analyzer
Reflection/RF Out
port or at the end of the connected extension cable.
4.
Press the
Continue
soft key to start calibration.
The calibration progress bar appears.
5.
Connect the
SHORT
connector of the CAL Kit directly to the
Cable and Antenna Analyzer
Reflection/RF Out
port or at the end of the connected extension cable.
6.
Press the
Continue
soft key to continue calibration.
Summary of Contents for CellAdvisor JD780A Series
Page 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Page 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Page 525: ......
Page 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...