![JDS Uniphase CellAdvisor JD780A Series User Manual Download Page 142](http://html1.mh-extra.com/html/jds-uniphase/celladvisor-jd780a-series/celladvisor-jd780a-series_user-manual_3602616142.webp)
JD780A Series Analyzers
142
Using GSM/GPRS/EDGE Signal Analyzer
NOTE
Measurement items such as IQ Origin Offset, C/I, EVM RMS/Peak and EVM 95
th
are applicable only to EDGE signal measurements.
Performing auto measurements
The Auto Measure function of the JD780A series allows a complete signal profiling covering RF
characterization and modulation quality parameters of up to 10 different carriers, particularly useful on
an overlay architecture where base stations are transmitting in different frequencies.
The Auto Measure can be easily executed either by selecting a menu in the instrument or by running a
programmed scenario in the PC-based application so that the instrument automatically configure and
perform tests on every aspect of all the carriers.
Setting limit for auto measure
You can set test limits for test item(s) in the auto measurement.
Procedure
1.
Press the
LIMIT
hot key.
2.
Press the
RF Test Limits
soft key, and then enable test limits as desired.
3.
Press the
PvsT Test Limits
soft key, and then enable test limits as desired.
4.
Press the
Modulation Test Limits
soft key, and then enable test limits as desired.
5.
Optional
. You can enable alarm sound that goes off if the measurement falls outside of the limit.
Toggle the
Beep
soft key between
On
and
Off
to enable or disable the beep sound.
6.
Optional
. Go to
SAVE/LOAD
> Save
, and then select
Limit
to save the limit settings.
See
“Using save” on page 31 for more information.
Setting measure setup
After configuring test parameters as described in the
“Configuring test parameters” on page 125, you
Summary of Contents for CellAdvisor JD780A Series
Page 52: ...JD780A Series Analyzers 52 Getting Started Figure 6 Example of Web remote on Chrome...
Page 254: ...JD780A Series Analyzers 254 Using LTE FDD Signal Analyzer 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 323: ...JD780A Series Analyzers Using LTE TDD Signal Analyzer 323 3 RS SINR 4 S SS RSSI 5 P SS Power...
Page 390: ...JD780A Series Analyzers 390 Using EV DO Signal Analyzer 3 Pilot 4 MAC 64 5 MAC 128...
Page 525: ......
Page 590: ...JD780A Series Analyzers 590 Appendix Figure 306 Append keywords window...