For cleaning or replacing following steps should be
done:
Switch off the power supply and wait for cooling
1.
down of the probe.
Turn away the filter lock (Pos 1) for pulling out the
2.
filter element.
‘Take-off the filter element (Pos 2) from the support
3.
tube of filter lock (Pos 1). Pull out filter element and
if applicable gaskets.
Replace the filter element (Pos 2) and/or the gaskets
4.
(applies only for sieve filter elements).
Remount the filter element (Pos 2) and if applicable
5.
the gaskets.
Screw on the filter element-screw stalwart.
6.
CAUTION
Tightening is radial. Tighten stalwart only!
NOTE
Any smell at the first time heat up is normal and is no
reason for a warranty claim.
New filter elements and sealings may influence the mea-
surement results. It is recommended to purge the gas
sampling probe diligently in heated condition.
Dimensions
10.
All dimensions in mm
Manual JPES
Für den Ersatz der Filterelemente sind folgende Schritte
vorzunehmen:
Elektrische Zuleitung abschalten und warten bis die
1.
Sonde abgekühlt ist.
Durch Drehen des Filterverschlusses (Pos 1) Filter-
2.
element entnehmen.
Filterelement (Pos. 2) vom Trägerelement (Pos 1)
3.
abziehen. Filterelement und ggf. Flach-Dichtungen
herausnehmen.
Filterelement (Pos. 2) und/oder Flachdichtungen
4.
(nur bei Siebgewebefilter) ersetzen.
Filterelement (Pos. 2) und ggf. Flachdichtungen
5.
montieren.
Filterelement-Verschraubung handfest anziehen.
6.
VORSICHT
Die Dichtung erfolgt radial. Nur handfest anziehen!
HINWEIS
Neue Filterelemente und Dichtungen können in den ers-
ten Stunden die Messergebnisse beeinflussen. Es wird
empfohlen, die Gasentnahmesonde in aufgeheiztem Zu-
stand ausreichend zu spülen.
Abmessungen
10.
Alle Abmessungen in mm
BA_DE_JPES_v1.5
–––––––––––––––