Technical data
3.
NOTE
If upgrade options are part of the sample probe the
information of the chapter 13 applies additionally and
has priority in case of doubt.
Subject to change without notice
*1...Mounting material is included
Technische Daten
3.
HINWEIS
Sind in der Gasentnahmesonde
Erweiterungsoptionen verbaut, gelten zusätzlich und im
Zweifelsfall vorrangig die Informationen in Kapitel 13
.
Technische Änderungen vorbehalten
*1...Montagematerial wird mitgeliefert
BA_DE_JES301E1_E1OS_v4.0
–––––––––––––––
[ 12 / 44 ]
–––––––––––––––––
Manual JES-301E1/V & JES-301E1OS/V
Betriebsdaten
Temperatur
max. 120 °C
Prozessdruck
max. 2 bara
Durchfluss
bis zu 600 Nl/h, applikations-
abhängig
Einbauwinkel
empfohlen 5°bis 15° aus der
Horizontalen fallend
Aufheizzeit
ca. 180 min
Einsatztemperatur
0 °C...55 °C
Zulässige
Umgebungstemperatur
-10 °C...55 °C
Einbaulage
Verdrehwinkel max. 45°
Durchfluss (Rückspülung)
ca. 4 Nm
3
/h @ 4 bar,
5 Nm
3
/h @ 6 bar
Rückspülmedium
Inertgas / Instrumentenluft
nur für Prozessabsperrventil (V)
Prozessabsperrventil
Kugelhahn, Temp. max. 200
°C @ 7 bar
Sitzmaterial
verstärktes PTFE
Konstruktion
Abmessungen über alles
168 x 415 x 316 mm BxHxT
Abmessungen
Klemmenkasten
115 x 115 x 64 mm BxHxT
Filterelement
2 µm Keramik oder 0,2 µm
Keramik oberflächenbe-
schichtet
Aktive Filteroberfläche
170 cm
2
Montageflansch
DN 65, PN 6; Form A nach
EN 1092-1; 1.4401
oder 2“ ANSI Flansch; 150
lbs., nach ASME B16.5
Messgas Eingang
G3/4“ Innengewinde
Messgas Ausgang
1/8“ NPT Innengewinde
Kalibriergasanschluss
(Option)
6 mm Rohrstutzen, 1.4401
Anschluss für Rückspülung
(Option)
6 mm Rohrstutzen, 1.4401
Mediumberührte Materialien
1.4404, Keramik, Viton
®
Heizelement
PTC, selbstlimitierend
Schutzart
Heizung: IP68
Gasentnahmesonde: IP43
Klemmenkasten IP66/67
Operational data
Temperature
max. 120 °C
Operating pressure
max. 2 bara
Flow rate
up to 600 Nl/h, depending on
application
Mounting angle
range 5° to 15° with respect
to the horizontal, sloping
down
Heat up time
approx. 180 min
Operating temperature range
0 °C...55 °C
Permissible
ambient temperature
-10 °C...55 °C
Mounting position
torsion angle max. 45°
Flow rate (back purge)
approx. 4 Nm
3
/h @ 4 bar,
5 Nm
3
/h @ 6 bar
Back purge medium
inert gas / instrument air
for process shut-off valve only (V)
Process shut-off valve
ball valve, temp. max. 200 °C @
7 bar
Seat material
reinforced PTFE
Construction
Dimensions over all
168 x 415 x 316 mm WxHxD
Dimensions
junction box
115 x 115 x 64 mm WxHxD
Filter element
2 µm ceramic or 0.2 µm
ceramic coated
Active filter surface
170 cm
2
Mounting flange
DN 65, PN6; form A accor-
ding to EN 1092-1; SS316
or 2“ ANSI 150 lbs; accor-
ding to ASME B16.5
Sample gas inlet
G3/4“ female thread
Sample gas outlet
1/8“ NPT female thread
Calibration gas port (option)
pipe 6 mm, SS316
Back purge port (option)
pipe 6 mm, SS316
Gas wetted materials
SS316L, ceramic, Viton
®
Heater element
PTC, self limiting
Protection class
heater: IP68
sample probe: IP43
junction box: IP66/67
Summary of Contents for JES-301E1/V
Page 32: ...BA_DE_JES301E1_E1OS_v4 0 32 44 Manual JES 301E1 V JES 301E1OS V...
Page 33: ...BA_DE_JES301E1_E1OS_v4 0 33 44 Manual JES 301E1 V JES 301E1OS V...
Page 34: ...BA_DE_JES301E1_E1OS_v4 0 34 44 Manual JES 301E1 V JES 301E1OS V...
Page 35: ...BA_DE_JES301E1_E1OS_v4 0 35 44 Manual JES 301E1 V JES 301E1OS V...
Page 38: ...BA_DE_JES301E1_E1OS_v4 0 38 44 Manual JES 301E1 V JES 301E1OS V...
Page 40: ...BA_DE_JES301E1_E1OS_v4 0 40 44 Manual JES 301E1 V JES 301E1OS V...
Page 41: ...BA_DE_JES301E1_E1OS_v4 0 41 44 Manual JES 301E1 V JES 301E1OS V...
Page 42: ...BA_DE_JES301E1_E1OS_v4 0 42 44 Manual JES 301E1 V JES 301E1OS V...