3. DATA ON MEASUREMENT BY PRINTER
SYSTOLIC : below 139 Hg
DIASTOLIC : below 89 Hg
MEANPRESS : 74~106 Hg
PULSE : 60~100beats/minute
PRP(Pressure Rate Product) : 7000~12000
Related with myocardial oxygen consumption
Note
The results of MEANPRESS and P.R.P are not displayed on LED, but printed out on
the paper.
Note
When pressure is high with the jammed air hose, message appears on the printer
as ERROR PRESSURE.
When the message is repeated, call for maintenance service.
When pressure is low as air leaks, message appears on the printer as ERROR
CUFF.
When the message is repeated, call for maintenance service.
When a testee moves or speaks while in testing, message appears on the printer
as ERROR MEASURE.
Try to re-test after for a while. If the message is repeated, call for maintenance
service.
A buzzer goes off once when error takes place, then the machine is ready for
testing
When the machine does not work altogether, turn the power off. And then make
sure if there is the fuse inserted and turn the power on and off a few times.
Results on printer paper
Normal Range
ID-NO
0001
DATE
02/15/2002
TIME
16:50
SYSTOLIC
122mmHg
DIASTOLIC
077mmHg
MEANPRESS
089mmHg
PULSE
080bpm
P.R.P
9760