background image

Manual for CD spectrometer

Subscribe to the outlook calendar (see below)

Go through the “Switch ON procedure” only 1-2 hours before using it!

Turn on the N

2

 flow on the “wall column”

. Only use the black wheel (not the pressure-reducing 

valve).

-  Turn on the “red box” (behind the box).

-  Adjust the N

2

 flow on the little black wheel to 12-20l/min.

Turn on the cooling water

N

2

 flow is 12-20l/min and the cooling water is on? If yes go ahead:

-  Tun on lamps of the power supply: POWER1, then POWER 2

-  Switch on the temperature controller (PTC-348MI) and press “Start”.

-  Start the computer and start the software on the desktop.

Sample Preparation:

1)  Prepare your sample in a volumetric flask, don’t fill it into a cuvette.

  Check on the UV/vis spectrometer, wheter you are in the right concentration range.

- Strong chromophores: c ≈ 10

-5

 M. If necessary, start diluting it after a first quick measurement.

- Weak chromophores: c ≈ 10

-4

 M. If necessary, start diluting it after a first quick measurement.

- GENERAL RULE: your absorbance band of interest should be at 0.1-0.8 E, ideally at around 

0.2-0.4 E.

2)  Fill your cuvette with pure solvent.

Start your measurement: (after the machine warmed up for 1-2 hours)

-  Choose “Setup”, then setup some parameters which you find in the submenus.

-  Insert your cuvette with pure solvent into the machine and record a Baseline. Save it.

-  Remove cuvette, empty it, blow it out with N

2

 and fill your sample solution. Put it into the      

machine.

-  Measure your sample.

-  Measure as many samples as your want, all will be automatically be corrected your baseline 

file.

Turn off the machine:

-  close the Specta Manager software shut down the computer

-  turn off the PTC-348WI turn off Power 2 switch 

-  turn off Power 1 switch 

-  purge instrument with N

2

 for 5 minutes after shutdown before turning off the nitrogen flow.

Reviews: