6

THE OPTITEX™ COVER

OptiTex is a durable hygiene cover with unique features contributing to the mattress’s 
pressure relieving properties.

The multi-stretch, breathable cover is made of a polyurethane (PU) material. The entire 
cover (bo

th upper and lower parts) can be cleaned with disinfectants or machine washed at 

maximum 95° C and tumble dried.

The OptiTex cover has an integrated waterproof cable holder that reduces the risk of bacteria 
spreading, keeps the electrical cord out of the way and protects the air hoses. 

Cells and hoses can easily be removed from the cover and rinsed clean with disinfectants or 
with warm water at maximum 70° C.

The Low Air Loss function

 allows air to escape from the cells whilst at the same time 

remaining waterproof. This microclimate management together with the bi-elastic surface 
layer reduces the risk of shear and other complications for the patient.

Fluid protection flap

 and waterproof zip on four sides, with welded seams.

Anti-slip underside

 which grips to bed base, in addition to four fastening straps.

Four transportation handles

 to enable easy moving.

APPLICATIONS

The OptiCell 4 PRO is an alternating pressure air mattress which is placed directly onto a bed 
base. It aids the prevention and treatment of pressure ulcers up to and including Category 4.

OptiCell 4 PRO has a height of 20cm and comes with a durable OptiTex cover and a CPR 
valve for emergency situations.

The supplied pump can provide both static and alternating pressure. It is easy to use 
and automatically adjusts to the patient’s weight, length, location, movements and 
position in the bed. The pump has four pre-programmed cycle times of 10, 15, 20 and 
30 minutes.

FIRE PROTECTION

The OptiCell 4 PRO mattress is tested and approved according to 

SS 8760001

EN 597-1

 and 

EN 597-2

.

Summary of Contents for OptiCell 4 PRO

Page 1: ...1 USER GUIDE OptiCell 4 PRO Alternating pressure air mattress...

Page 2: ...roduct 8 Receiving and unpacking 8 Storage 8 Automatic adjustment and alternating pressure 9 Static mode 9 Low pressure alarm 9 Air hose connection 10 Hand Check test 11 Inspection and troubleshooting...

Page 3: ...e CE marking signifies that OptiCell AB has the manufacturer s responsibility to ensure that this product meets the essential requirements of relevant European legislation This includes the provision...

Page 4: ...ment of the intended usage environment and then use the right product Please contact us if you have any questions In order to avoid damage the mattress should not come into contact with heat sources o...

Page 5: ...use maintenance and improved safety It is very important to read this guide carefully before using the product Care personnel must be given access to this user guide and also be informed of the risks...

Page 6: ...the bi elastic surface layer reduces the risk of shear and other complications for the patient Fluid protection flap and waterproof zip on four sides with welded seams Anti slip underside which grips...

Page 7: ...ell 4 PRO air mattress is connected to the pump by three air hoses The red air hose positioned in the middle is used to fill the head cells situated under the patient s head It also fills the lower ch...

Page 8: ...n mind that a tall patient can weigh the same as a short patient but may need a different pressure This must always be carefully observed for each patient The biggest risk for the patient is to bottom...

Page 9: ...not want to be disturbed by an alternating pressure Press the Care Static button position E on page 14 to select either Care mode or Static mode Care also called Auto firm mode provides maximum pressu...

Page 10: ...connector if the air hose is disconnected from the pump To minimise the risk of air leakage during hose connection check That the two hooks are not damaged and their clips are intact That the black o...

Page 11: ...t is easier for the staff but it may give an incorrect test result Fold away the sheets and unzip the mattress cover to access the cells Put one hand between two cells with the patient lying on the ma...

Page 12: ...r hoses for signs of damage or abnormal wear Check the air hose connector see page 10 INSPECTION OFTHE OPTITEX COVER Regularly check the cover for visible signs of damage or wear such as tears cracks...

Page 13: ...d at a minimum of 70 C for at least 10 minutes with unscented detergent according to standard BS EN 14065 2016 Textiles Laundry Processed Textiles Biocontamination Control System The OptiTex cover can...

Page 14: ...ng on the mattress is affected by the comfort mode chosen Press this button to cycle through modes ALARM If the alarm sounds press this button to mute and reset The adjacent Leaking lamp will light up...

Page 15: ...out condensation Storage 10 90 without condensation Safety standard CE marked CARE STATIC Press this button to select Care auto firm mode or Static mode When the adjacent Care lamp is lit the pump pro...

Page 16: ...posed of at a recycling station and sorted as Combustible waste For more information contact J rven HealthCare J rven Health Care takes environmental responsibility seriously We strive to always use m...

Page 17: ...COMPLAINTS Any complaints regarding the product must be notified in writing to J rven Health Care The return of products must be done in accordance with our instructions Returning a product before a c...

Page 18: ...essment is made to identify the specific needs of the patient regarding pressure relief and must be conducted by staff with adequate knowledge and education Based on the patient s assessed status the...

Page 19: ...19 YOUR NOTES If you require any further information or assistance please get in touch with us We will be happy to help you...