2
Upper Arm Type
Blood Pressure Monitor
Model Number: B07
--------------------------------------------------------------------------
USER MANUAL
Page 1: ...nity Guanlan Street Longhua District Shenzhen 518100 PEOPLE S REPUBLIC OF CHINA Website www jiemeirui com Telephone 0086 755 85292057 Authorized European Representative Company Name Shanghai Internati...
Page 2: ...2 Upper Arm Type Blood Pressure Monitor Model Number B07 USER MANUAL...
Page 3: ...essure and its Measurement 5 3 Components of your blood pressure monitor 7 4 Using your Monitor for the First Time 8 5 Measurement Procedure 9 6 Care and Maintenance 16 7 Warranty 17 8 Certifications...
Page 4: ...d to replace regular medical checkups It is recommended that your physician review your procedure for using this device Blood pressure readings obtained by this device should be verified before prescr...
Page 5: ...humidity conditions outside the limits stated in the specifications section of this manual Warning The separate ac adapter which is intended to connect USB interface of Blood Pressure Monitor has not...
Page 6: ...g the frequency of heart pacemakers Attention In cases of irregular heartbeat measurements made with this instrument should only be evaluated after consultation with your doctor Note To obtain the gre...
Page 7: ...ecessary If you have blood pressure values that are too low i e systolic values under 105mmHg and or diastolic values under 60 mmHg consult your physician Even with normal blood pressure values a regu...
Page 8: ...o important organs heart brain muscles can result from arteriosclerosis Furthermore the heart will become structurally damaged with increased blood pressure values 3 There are many different causes of...
Page 9: ...Features of Model B07 1 Talking function 2 Double users 2 x 120 sets memory 3 Cuff self checking function 4 Irregular heartbeat checking 5 Average value function 6 Low battery display 7 WHO function 8...
Page 10: ...ess power button to confirm Setting Volume When display with SP is flashing press memory button to switch volume 1 volume 2 volume 3 or OFF Press power button to confirm Record Delete When you checkin...
Page 11: ...the accumulated blood to flow away If you decide to take your Averaging Mode measurement again be sure to wait at least one minute beforehand 5 3 Fitting the Cuff Please refer to picture 06 a The cuff...
Page 12: ...ing make sure the palm of hand is upturned Make sure arm is in correct position to avoid body movement Sit still and do not talk or move during the measurement After the cuff has been appropriately po...
Page 13: ...emove the cuff and wait 2 3 minutes Reapply the cuff and take another measurement Recommended Use Methods 1 recommendation that the PATIENT relax as much as possible and not talk during the measuremen...
Page 14: ...If the symbol appears more frequently e g several times per week on measurements performed daily or if it suddenly appears more often than usual we recommend the patient to seek medical advice The ins...
Page 15: ...in 15s Tie the cuff correctly and make sure the air plug is properly inserted in the unit Er 5 The air tube is crimped Correct it and make the measurement again Er 6 The sensor is sensing great fluctu...
Page 16: ...od Pressure Monitor B07 and its use Authorized Representative in the European Community CE Mark conforms to essential requirements of REGULATION EU 2017 745 Date of manufacture Manufacturer SN Specifi...
Page 17: ...it is necessary to interrupt a blood pressure measurement for any reason e g the patient feels unwell the Power button can be pressed at any time The device then immediately lowers the cuff pressure...
Page 18: ...follow manufacturer s instructions carefully Change the battery may result in unacceptable risks the operator should pay attention to the following Do not insert batteries with their polarities incorr...
Page 19: ...sher or submerged in water d Handle the tube carefully Do not pull on it Do not allow the tubing to kink and keep it away from sharp edges e Do not drop the monitor or treat it roughly in any way Avoi...
Page 20: ...Pa 106Kpa Measuring method Oscillometric Operating method Continuous operation Pressure sensor Resistive Measuring range DIA 30 195mmHg SYS 60 250mmHg Pulse 40 to 199 per minute Cuff pressure display...
Page 21: ...ice is intended for use in the electromagnetic environment specified below The customer of the user of the device should assure that it is used in such an environment Emission test Compliance Electrom...
Page 22: ...EC 61000 4 5 1 kV line s to line s 2 kV line s to earth Not applicable Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage v...
Page 23: ...ccording to the transmitter manufacturer and d is the recommended separation distance Field strengths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the...
Page 24: ...quency of transmitter m 150 KHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 7 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum outp...
Page 25: ...sing higher IMMUNITY TEST LEVELS that are appropriate for the reduced minimum separation distance Minimum separation distances for higher IMMUNITY TEST LEVELS shall be calculated using the following e...
Page 26: ...26...