JADEVER SCALE CO., LTD
JWE service manual
15
Description of Parameter Values
1)
Calibration See “7. Calibration”
2)
Set the active weight unit. The following setting are available:
kg, g, lb, lb-oz,
斤
,
台斤
,
港斤
,pcs and % .
(Enable);
(disable)
3)
Set the init weight unit. The following setting are available:
kg, g, lb, lb-oz,
斤
,
台斤
,
港斤
,pcs and %.
(Enable);
(Disable)
4)
Set the filtering level in which the stable indication turns on .The higher the setting,
the slower stabilization time. Options are Fil 1,Fil 2,Fil 4 and Fil 8.
5)
Set the period of inactivity before the scale automatically turns off. Options are
Aut.NO=non power-off,Aut.5, Aut.10, Aut.30 and Aut.60(minutes).
6)
Set the activation mode of backlight. Options are lit.Aut=Auto on with items greater
than 9d placed on the weighing pan, lit.ON = Backlight on and lit.OFF=Backlight off.
7)
Set zero display range. Options are Zero.0(one division) , Zero.1(2 divisions),
Zero.2(3 divisions), Zero.3(4 divisions), Zero.4(5 divisions) and Zero.5(6 divisions) .
8)
Set RS-232 Serial Transmission Rate. Options are bAu.96(9600), bAu.48(4800) and
bAu.24(2400)
9)
Set Print mode .Options are Prt.Pr(manual print), Prt.St(Stable print) and
Prt.Co(Continuous print)
Note:
In the continuous print mode, computer will be automatic chosen as the external device.
10)
→
Set external Interface. Options are PC, SH, ZEBRA,TP, AX, EZ2-S and
tdp-643.
11)
Set Weight checking memory.
(Enable);
(Disable )
12)
Set Weight checking buzzer beep. Options are beep.Un, beep.IN, beep.NO and
beep.LO.