only 1-D bar code is supported.
4.3 System Set
Press to initiate system setup
1)
Auto-Off:
Sets the period of inactivity before the scale automatically turns
off.
2)
Sound:
The scale has been installed button sound and weighing speaker.
Please connect to audio device, if you wish to use this function.
3)
Manual Update:
Use the U disk to copy the new program, then insert the U
disk to the USB port which has the marking of “dev”. The system can be
updated by simply pressing Manual Update button.
4)
Theme:
Press to select your favor theme: blue style or default.
5)
Language
:
Simplified Chinese/Traditional Chinese/English/Spanish
6)
TIME:
Set time
7)
AutoSleep:
Sets the period of inactivity before the scale automatically turns
to standby mode.
Note: The way to wake up the scale
◎
Press key
Sleep/Wake
◎
Click the screen
◎
Under weighing mode, click the weighing pan or the screen
4.4 CAL – Calibration
:
For best results calibrate the scale at regular
intervals. This is especially important if the scale is in use for prolonged
periods.
Calibration method I
Arbitrary value
: After the zero point calibration finished, input any
calibration value to continue the calibration procedure.
Calibration method II
Cal Setup
: Input the calibration points. It will automatically work out each
calibration point value
Multi-point:
After setup the calibration point, conduct the calibration
procedure.
Note: When the unit is “Ton”, the capacity and calibration value are set by
unit “kg”
4.5 Set Weight
1) Weight CK-- Check weighing:
Use this mode to compare the weight of an
item to Lower and Upper limits.
High
The weight on the weighing pan is greater than the upper limit
OK
The weight on the weighing pan is between upper and lower limits.
Low
The weight on the weighing pan is smaller than lower limit.