IDS701-CTRUCK User Manual
Page 11 of 21
press
press
or
press
press
or
press
[C06 0]
[C06 1]
[SPAN ]
[0100.00]
[0080.00]
[CAL 9]
……
[0080.00]
[CALEnd]
option:
0=Non-load calibration
1= load calibration
Basic on max capacity setting,
add suitable weight on scale.
close to the max capacity,
heavier than 10% max at least.
For example: the weight is 80kg
As bellows:
enter loading calibration, count
down over, indicator shows
loaded
weight
,
loading
calibration finish.
If you want to set application
Function parameter. Press
“PRINT” if you want to exit press
“TOTAL”
8
press
press
press
or
[C07 ]
[07 0]
[07 1]
Default parameters setting
option:0=non-restore default
parameters
1=restore default parameters
Note:
after
the
above
parameters setting finish, please
do not set default parameters
often, avoid the original setting
parameters lost.