Chapter 5
– Setup Menu
58
FIGURE 51 -
R
G
B
MEASUREMENT SCREEN
2. The instrument will make continuous measurements until the
Abort
key is touched. Adjust the primaries
of the display until the vertical indicator bars for
R
,
G
and
B
are all at the center of the respective
horizontal colored bars and the % deviation (values to the right of the bars) reads 0.00.
Note that updated
Luminance, CCT
and
chromaticity
values are continuously displayed during the
calibration sequence.
3. Touch
Abort
to
stop the measurement.
4. Touch
Back
to exit this screen and return to the Measurement Setup Screen at the completion of the
calibration.
C
OLOR
R
ENDERING
I
NDEX
(CRI)
–
O
PTIONAL FOR
PR-7XX.
Color Rendering Index is a method of testing luminaires or other light sources for their capability to render the
color of an object being illuminated by that source with respect to a standard light source
– does the object look
the same under both sources. Fourteen reference colors are used to determine the similarities between the test
light and the standard with the first eight colors used to calculate the average CRI
– see Figure 52.
FIGURE 52 - COLORS USED TO DETERMINE AVERAGE CRI
Summary of Contents for Photo Research SpectraScan PR-7 Series
Page 2: ......
Page 10: ...PR 7XX User s Manual vi...
Page 32: ......
Page 84: ......
Page 91: ...Chapter 7 File Settings 87 FIGURE 73 ILLUMINANCE LUMINANCE CONCEPT...
Page 92: ......
Page 102: ......
Page 108: ......
Page 117: ...Chapter 12 Connectivity 113...
Page 118: ......