Chapter 9
– How to Make an Illuminance / Irradiance
Measurement
97
U
SING THE
ICC-730
I
NTEGRATING
S
PHERE
FIGURE 81 - ICC-730 INTEGRATING SPHERE
The ICC-730
is a 3” (76.2 mm) integrating sphere with a 1 inch (25.4 mm) input port. A typical application is
determining the illuminance of point sources such as lamps.
M
EASURING
P
ROCEDURE
1. From the
Instrument Setup
menu, touch the ▲▼icons beneath
Primary Accessory
to scroll to ICC-730.
Make any other setup changes (exposure, sensitivity etc.) at this time (see Chapter - 5 - Setup Menu for
more details).
2. Remove the current
Primary Accessory
from the instrument by turning the accessory counterclockwise.
3. Install the ICC-730 by turning clockwise.
Hand tighten only. Do not over-tighten!
4. Set the front edge of the entrance port to the desired distance from the source. For the best results, the
measuring port should be set so that it is perpendicular to the source.
5.
Press the
MEASURE
button to begin a measurement.