8.7 Items to inspect during annual maintenance
1
Exterior
The exterior of sterilizer should be tidy and mustn’t have the disfigurements,
such as deflection, hollowness, collision, nick, sharp edge, and so on.
2
Cover plate
The cover plate should be assured to disassemble easily in order to repair
the equipment.
3
Digit and letter
The digit and letter in screen should be legible.
4
Electroplate components
The Electroplate should accord with YYOO76-1992 class 2, which for the
request of aspect.
5
Printer components
The Printer components should accord with YY1055-1999 the class 2 which
for the request of aspect.
6
Door safe lock
On the normal condition, if the sterilizer door hasn’t been locked tightly, the
program can not start.
7
Chamber Pressure
The sterilizer should ensure that the door can’t be opened when chamber
pressure is greater than 0.27Mpa.
8
Safety Valve
The sterilizer must install a safety valve, safety valve opening pressure
0.27Mpa ± 0.01Mpa, and automatically open when reaching the set value.
9
Sterilizing Program
Sterilizer should have the pre-established program about 121ºC and 135ºC,
dressing and instruments.
10
Controlling system
The control system in sterilizer should limit the steam which in the chamber
be controlled at the highest average temperature in ±3ºC of pre-establish
station. And ensure the temperature value accords with the pressure control
-
ling value.
11
Timing control
Able to timing control of the sterilization and drying, and the error should
not be greater than 10% of the preset value.
12
Button and Switch
Buttons and switches should be flexible and reliable on the sterilizer.
13
Indicator and display
The indicators and displays of sterilizer should show the states of every
sterilizing procedure exactly. Under the normal situation, sterilizer should
indicate:
a) Chamber temperature
b) Chamber pressure
c) Sterilizer working state
d) Water level state
e) State of door
8.6 List of alarm codes
Error Alarm List
CODE No.
Error Information
Abnormal exit
01
Temperature sensor fault (steam generator)
02
Temperature sensor fault (out chamber)
03
Temperature sensor fault (inner chamber)
04
Sterilization fail
05
Pressure not exhausted
06
Door open in the cycle
07
Operation over time
08
Over pressure
09
Over temperature
10
Pressure and temperature do not match
11
Vacuum process fail
98
Power fail