24
Displayed temperature is
incorrect.
[Note: Types ‘K’ & ‘J’
The controller has not warmed-up.
The display temperature reads low when the controller is first
turned on, but will self-correct as it warms up. The controller can
be used immediately since it will warm up during the initial stages
of heating.
Thermocouples display negative
temperatures, but are not
calibrated for them]
Wrong type of thermocouple is
plugged into controller.
Thermocouples are color coded. Thermocouple plug and
thermocouple receptacle must be the same color (see Section 3.7).
Corroded thermocouple
connections.
Clean plug on thermocouple and receptacle on controller with
sandpaper or steel wool.
Corroded thermocouple.
If the temperature-measuring end of the thermocouple is corroded,
discard thermocouple.
Temperature display offset needed.
To enter a controller display offset:
1.
Turn on controller. Allow to warm up for 30 minutes.
2.
Record displayed temperature.
3.
Press both the
▲
and
▼
keys on the front of the temperature
meter for the channel you want to change until “tunE”
appears, then let go of the keys.
4. Press
the
▼
key until “LEVL” appears.
5.
First hold in the “*” key, then while holding in the “*” key
press the
▲
key until “3” is showing in the display, then let
go of all keys.
6. Press
the
▲
key until “ZEro” is showing in the display.
7.
Note the current display offset (this is the number blinking in
the display).
8.
Calculate the new offset temperature using the equation:
New Current
Display = display offset - Displayed + Correct
Offset blinking in display temperature temperature
9.
First hold in the “*” key, then while holding in the “*” key
press the
▲
or
▼
keys until the new offset temperature is
showing, then let go of all the keys.
10. Press
the
▲
and
▼
keys until the temperature is displayed.