M-2000® Hardware Manual
©2023 J.A. Woollam Co.
43
Figure 3-39. Verifying control type.
8.
The display will now ask “TCP OK: xxx:xxx:xxx:xxx?” If it is correct use the
down arrow key to select yes.
Figure 3-40. Verifying TCP/IP address.
9.
The display will now ask “Mask OK: xxx:xxx:xxx:xxx?” If it is correct use
the down arrow key to select yes.
Figure 3-41. Verifying Mask address.
10.
The EC has now been setup for the TCP/IP address.
Note:
The default TCP/IP address set at the factory is 192.168.000.002.
The default Mask is 255.255.255.000. In order to establish proper
communications between EC and the operator’s PC, the network
card in the PC must be set correctly also.
TCP/IP address in the hardware.cnf file
1.
Using Notepad, open the hardware.cnf file located in the
C:\CompleteEASE\cnf directory.
2.
The line under the [Hardware] section should say
‘Remote=TCP 192.168.0.2 300 0’
3.
The first four numbers (192.168.0.2) need to be the same as the TCP/IP
address on the EC box.
4.
If a change is made, save the file, and changes will be in effect the next
time CompleteEASE is started.
3.3
Operator Computer Setup
If the operator computer was not sent with the instrument, there are some
additional steps required.
Summary of Contents for M-2000
Page 2: ...2 2023 J A Woollam Co M 2000 Hardware Manual...
Page 4: ......
Page 9: ...M 2000 Hardware Manual 2023 J A Woollam Co ix...
Page 10: ......
Page 50: ......
Page 152: ...152 2023 J A Woollam Co M 2000 Hardware Manual...
Page 172: ......
Page 175: ...M 2000 Hardware Manual 2023 J A Woollam Co 175 M 2000V I M 2000X I...
Page 177: ...M 2000 Hardware Manual 2023 J A Woollam Co 177...