M-2000® Hardware Manual
©2023 J.A. Woollam Co.
169
12.4
Software Problems and
Error Messages
SYMPTOM
CAUSE
SOLUTION
CompleteEASE
does not
initialize.
Cables are not
plugged into
computer or EC.
Ensure all cables are inserted and fastened.
Recent hardware
changes to
computer such as
a network card.
Remove recent additions to computer. Then,
correct hardware settings of new cards such
that they don’t conflict.
Table 12-2. Software problems and error messages.
12.5
Calibration Problems
SYMPTOM
CAUSE
SOLUTION
Calibration fits are
not acceptable.
Sample is not aligned
properly.
Re-align the sample.
Signal intensity is weak
on some channels due
to sample reflection
properties. Detector(s)
Overloaded.
Use the standard SiO2 on Si
reference sample supplied with the
instrument for calibration. Close the
respective iris or choose a higher
density ND filter (if equipped).
Software installed
on a new computer,
and results are not
correct.
Calibration files not
installed on the new
computer
Copy cnf files from the previous
computer into the
C:\CompleteEASE\cnf directory of
the new computer
Table 12-3. Calibration problems.
12.6
Signal Screen Problems
SYMPTOM
CAUSE
SOLUTION
Signal screen shows flat
lines on all channels.
Cables not connected
properly. Electronic
scanning problems.
Make proper cable
connections. Call J.A.
Woollam Co. for
instructions.
Table 12-4. Signal screen problems.
Summary of Contents for M-2000
Page 2: ...2 2023 J A Woollam Co M 2000 Hardware Manual...
Page 4: ......
Page 9: ...M 2000 Hardware Manual 2023 J A Woollam Co ix...
Page 10: ......
Page 50: ......
Page 152: ...152 2023 J A Woollam Co M 2000 Hardware Manual...
Page 172: ......
Page 175: ...M 2000 Hardware Manual 2023 J A Woollam Co 175 M 2000V I M 2000X I...
Page 177: ...M 2000 Hardware Manual 2023 J A Woollam Co 177...