2.6
Connecting the Components
Here we will describe how to connect the ModBox system to your application.
Step Action
1.
You should prepare at least 2 optical fiber patch cords (with FC or SC connectors
depending purchased option) to use the ModBox system with the embedded
reference photo receiver (if you purchased this option). Make sure that the fiber
connector end facets are clean before using them with the ModBox.
2.
You should also prepare an electrical PRBS data signal to be fed in the system,
you will need a 2.92 mm or 1.85 mm RF connector cable to connect with the
ModBox RF input.
3.
Connect one of the laser optical output fiber connectors to a power meter using
an optical fiber patch cord (delivered with the ModBox).