iWorX GA-200 Series Technical Note Download Page 7

6) Click 

Stop

.

7) Select 

Save As

 in the 

File

 menu, type a name for the file. Choose a destination 

on the computer in which to save the file. Designate the file type as *.iwxdata. 
Click 

Save

.

To Convert the Units on Gas Concentration Channels:

1) Use the 

Display Time

 icons to adjust the display time of the Main window to 

show the complete calibration data on the Main window at the same time. The 
required data can also be selected by:

• 

Placing the cursors on either side of data required, and;

• 

Clicking the Zoom between Cursors button on the LabScribe toolbar to 

expand the entire segment of data to the width of the Main window.

2)

Click the 2-Cursor icon on the LabScribe toolbar so that two cursors appear on 
the Main window. Place one cursor on the section of data recorded when gas 
analyzer was collecting a sample of room air and the second cursor on the 
section of data recorded when the second sample was collected.

The LabScribe toolbar.

To convert the voltages at the positions of the cursors to concentrations 
using the Advanced Units Conversion dialogue window:

1) To convert the voltages on the 

Expired CO2 Concentration (%)

 channel, click 

on the arrow to the left of the channel title to open the channel menu. Select 

Units

 from the channel menu, and select 

Advanced 

from the 

Units

 submenu.

The voltage outputs of the two sensors in the GA-200 gas analyzer, carbon dioxide on  
the top and oxygen on the bottom. Other recording windows have been minimized to 
show detail.

GA-200 Gas Analyzer 

iWorx Systems, Inc.

www.iworx.com

Summary of Contents for GA-200 Series

Page 1: ...oxide in a gas a sample is pumped from the input port on the front panel of the GA 200 through the sample cell and out an exhaust port on the back of the unit Gas samples are drawn into the unit throu...

Page 2: ...to a computer through a serial RS 232 data stream The default data format is 9600 baud 8 bit and no parity with alternate baud rates of 2400 4800 19200 and 38400 The signals present are RXD TXD and g...

Page 3: ...the 20 mA current output is set to 100 Socket 13 is the Ground for the both current outputs Socket 14 is the oxygen current output Socket 15 is the carbon dioxide current output GA 200B Rear Panel GA...

Page 4: ...the surface of a detector The output of the sensor is inversely proportional to the concentration of oxygen in the sample because the amount of light reaching the detector decreases as the concentrati...

Page 5: ...en the output of the carbon dioxide sensor which is located on the terminal block on the rear panel of the analyzer and the BNC input of Channel 1 on the IWX 214 Red wire into socket 15 and black wire...

Page 6: ...lve on the regulator of the cylinder providing the second gas sample Adjust the flow rate to low Make sure the flow rate from the gas cylinder is greater than the flow rate on the gas analyzer 2 While...

Page 7: ...cribe toolbar so that two cursors appear on the Main window Place one cursor on the section of data recorded when gas analyzer was collecting a sample of room air and the second cursor on the section...

Page 8: ...the GA 200 gas analyzer 5 Notice that the voltages from the positions between the cursors are automatically entered into the value equations Enter the two concentrations of carbon dioxide measured fr...

Page 9: ...be disinfected An external barrier filter on the inlet port must be used to prevent contamination of the sample circuit 4 If the GA 200 is used to collect samples of inspired and expired air during b...

Page 10: ...zer 12VDC 1 5A Requirements of the External Power Supply 95 250VAC 47 63 Hz Dimensions 190 x 76 x 280mm WxHxD Weight Analyzer 2 3kg Power Supply 0 7kg Storing Before storing the GA 200 connect an exte...

Reviews: