Chapter 4 Measurement
18
Chapter 4 Measurement
4.1 Preparation for Measurement
(1) Have this instrument, a monitoring device, and a target to be
measured ready.
*: A monitoring device means an oscilloscope, DVM, recorder, and
other devices.
(2) Place four AA dry batteries in the main unit.
<For batteries>
a) To open the case, push downward on the lid of the battery case
on the back of the main unit (refer to
in the appearance in
Section 2.1).
b) Place four AA dry batteries in the battery case.
Note) Do not place them in the wrong orientation.
c) Close the lid of the battery case.
(3) Connect the accessory BNC cable to this instrument.
a) Connect one end of the BNC cable to the output part on the top of
this instrument (refer to
in the appearance in Section 2.1).
b) Attach the other end of the BNC cable to the monitoring device.
Caution
・
To use this instrument, connect it to the ground
(output side).
Ground GND terminals of monitoring devices, which are
the connection destination of the output terminal.
Do not energize a target to be measured with nothing
connected to the output terminal. This may be a cause of
an electrical shock.