CAUTION
!
About the caution marks and the voltages that display in the panel and the
patch panel of the test fixture
Missing proper wiring and/or wrong connection may conduct wrong measurement. Also these
connections may cause of damage on DUT and/or malfunction of the unit.
Case1: Wrong connection between Collector terminal and Base terminal;
DUT may be damaged while extra high voltage can be supplied on Base terminal.
Case2: NOT connecting Collector output terminal, Base output terminal and/or Emitter terminal
and connecting Collector sense terminal, Base sense terminal and/or Emitter sense terminal
respectively;
It may cause for burn of SENSE RESISTERs inside of the unit since the sense resisters
connected to outputs respectively for Kelvin sensing and may be supplied current exceeding
each specification.
Case3: Connecting HIGH VOLTAGE terminal and HIGH CURRENT terminal of collector output at
the same time on DUT;
It may be cause for fire and/or malfunction due to high voltage applied to HIGH CURRENT
terminal.