Capter 3
Function Setting and Basic Operation
Curve Tracer CS-3000 Series Instruction Manual
3-33
3.7.4 VOLTS key
3.7.4.1
VOLTS key and display menu
Pressing VOLTS key
④
allows "MAX PEAK VOLTAGE" menu to appear.
High Voltage mode
High Current mode
3.7.4.2
Submenus and functions
If the maximum voltage of 3 kV is selected, "
3 kV"
on COLLECTOR SUPPLY display menu changes to red.
3.7.4.3
Relationship with other menus
The voltage setting value and the maximum peak power setting value determine the value of
the current limitation resistance internally connected to the collector output in series. See Table
3.7 POWER WATTS/VOLTAGE and the current limitation resistance in Section 3.7.
If this setting is changed, VARIABLE setting value is cleared to 0%. See Section 3.7.1.2
"Initializing VARIABLE setting value."
If the maximum peak power is set to 390 W and the configuration is set to
"Open/Collector/Common" (i.e.: the collector output is connected to the base), 3 kV cannot be
selected.
Key
High Voltage mode
High Current mode
F1
Sets the maximum voltage to be applied to a
measured device (3kV, 300V, or 30V).
40V fixed
F2
F2 to F6:(Not used)
F3
F4 to F6 (Not used)
F1
F2
F3
F4
F5
F6
* CS-3200 and CS-3300 include
HIGH CURRENT.
Summary of Contents for CS-3000 Series