Part No. 2718-55-16

OPERATION MANUAL

© 2014 IVY Biomedical Systems Inc.   All Rights Reserved. 

Part No. 2718-24-16

Cardiac Trigger Monitor 

Model 3150-B  

Designed exclusively to operate with 

GE Healthcare CT Scanners

GE Part Number: 5304770

 

Summary of Contents for 3150-B

Page 1: ...16 OPERATION MANUAL 2014 IVY Biomedical Systems Inc All Rights Reserved Part No 2718 24 16 Cardiac Trigger Monitor Model 3150 B Designed exclusively to operate with GE Healthcare CT Scanners GE Part...

Page 2: ...by IVY Biomedical Systems Inc trained personnel The product must not be altered without the prior written approval of IVY Biomedical Systems Inc Quality Assurance Department The user of this Product s...

Page 3: ...This page is intentionally left blank...

Page 4: ...10 Programmable Keys 11 Menu Structure 12 Display 13 Alarm Messages 14 Rear Panel 14 Fuse Ratings 15 MONITOR SETUP 16 Set up the instrument for operation 16 Change Mains Voltage 16 Set the Language 1...

Page 5: ...CG and Impedance Data Transfer using the USB Port 28 USB Port 28 RECORDER OPERATION 29 Changing Paper 29 Recorder Modes 30 Recorder Speed 31 Example Printout 31 ALARM MESSAGES 32 Low Signal message 32...

Page 6: ...examination by Ivy Biomedical Systems Inc discloses such product s or component part s to have been defective then Ivy s obligation is limited at Ivy s option to repair or replacement When a product...

Page 7: ...WARRANTY iv Model 3150 B Operation Manual This page is intentionally left blank...

Page 8: ...ype is used in text to refer to the labeling on user controls Special brackets surround menu selections used with the programmable keys Manufacturer s Responsibility The manufacturer of this equipment...

Page 9: ...about the integrity of the protective ground conductor arrangement do not operate the monitor until the AC power source protective conductor is fully functional WARNING Do not place the monitor in any...

Page 10: ...the alarms are set to off neither visual nor audio alarms will be present MRI The Model 3150 B should not be used within the magnetic field during Magnetic Resonance Imaging Pacemakers Rate meters mig...

Page 11: ...is 0 5 mV AAMI EC 13 3 2 6 1 The use of the Model 3150 B below the above amplitude value may cause inaccurate results Guidance and manufacturer s declaration Electromagnetic emissions The Model 3150 B...

Page 12: ...power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality sho...

Page 13: ...ths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of the equipm...

Page 14: ...IEC 60601 1 or IEC 950 with configuration to IEC 60601 1 1 j Type CF applied part Defibrillator proof Fuse type rating Output signal _______ ON Input signal Stand By STBY a Alternate Current AC Protec...

Page 15: ...he monitor and the CT console for the transfer of ECG data and trigger timing data and the receipt of patient identification information This function will only operate when the Model 3150 B is electr...

Page 16: ...art Defibrillator proof ECG Degree of protection against harmful ingress of water Ordinary equipment IPX0 per IEC 60529 Methods of Maintenance and Cleaning See page 35 Degree of safety of application...

Page 17: ...ff the alarms This avoids the problem of turning off the alarms and forgetting to turn them back on Press this key again to return the alarms to normal before the two minutes have expired Pressing ALA...

Page 18: ...rammable Keys Displayed above each programmable key is either a menu item or a function Pressing a programmable key will display other menu levels or activate an appropriate function Menu functions ar...

Page 19: ...R DET ECT E NABLED and DIS ABLED NOT CH FILT ER ON and OFF IMPEDANCE ENABLED and DISABLED P LOCK E NABLED and DIS ABLED Low 10 to 245 bpm High 15 to 250 bpm ECG Menu Limits M enu T est OPTIONSMENU Mo...

Page 20: ...of the status of the connection between the Model 3150 B and the CT Scanner XRAY On Off Indicates that the CT Scanner X Ray is ON or the CT Scanner X Ray is OFF The XRAY On Off indicator is located i...

Page 21: ...is is to ensure the connection complies with leakage current requirements of one of the following applicable standards UL60601 1 CAN CSA C22 2 No 601 1 M90 IEC 60601 2 25 and CE MDD 93 42 EEC The maxi...

Page 22: ...ce that the safety certification of the ACCESSORY has been performed in accordance to the appropriate IEC 60601 1 and or IEC 60601 1 1 harmonized national standard Model 3150 B Rear Panel Fuse Ratings...

Page 23: ...g the ON key 3 Press the LANGUAGE key to set the desire language The language choices are English Spanish French German Italian Portuguese Swedish Danish Dutch Norwegian and Finnish 4 Turn monitor off...

Page 24: ...the ON key Setting Initial Default Auto Impedance Checking OFF Initial Language Setting English ECG Size 10mm Lead II Trigger output mark ON ECG Notch filter ON Impedance Enabled Impedance Threshold...

Page 25: ...amplitude typically Lead II x The proper placement of the ECG electrodes The ECG electrodes may need to be repositioned x The ECG electrodes still have moist conductive gel Polarity Lock P LOCK With s...

Page 26: ...ductors at the cable end Unprotected lead wires and cables may pose an unreasonable risk of adverse health consequences or death Leakage current is limited internally by this monitor to less than 10 A...

Page 27: ...ver chloride ECG Electrodes or equivalent For best ECG performance use ECG Electrodes supplied by Ivy Biomedical Systems see Accessories Use the following procedure for ECG monitoring 1 Prepare each e...

Page 28: ...000 50k The use of the wrong type of ECG electrodes improper application or poor skin preparation can increase the electrode impedance value causing an imbalance between the leads which can allow nois...

Page 29: ...litude Size Use the following procedure to adjust the amplitude size of the displayed ECG waveform 1 Press the SETUP key from the main menu The following menu appears STBY ON ON ECG 150 HR I FILT 60 1...

Page 30: ...programmable key to adjust the ECG waveform amplitude 4 Press EXIT to return to the main menu Lead Selection 1 Press the SETUP key from the main menu 2 Press the first programmable key ECG once to sel...

Page 31: ...G electrodes still have moist conductive gel ECG Notch Filter Use the following procedure to activate the Notch filter 1 Press the SETUP key from the main menu 2 Press the ECG key and select the OPTIO...

Page 32: ...EXIT to return to the main menu Alarm Type Default Limit Heart Rate Low 30 Heart Rate High 160 Pacemaker Follow the next procedure to activate or deactivate the pacemaker detection function 1 Press t...

Page 33: ...or the CT Scanner X Ray is OFF 1 Large symbols in the upper right hand corner of the display provide the operator with a visual indication of the status of the connection between the monitor and the...

Page 34: ...he top of the display above the heart rate The same Patient ID is also stored for transmission back to the CT console if a subsequent Data Recall operation is performed In addition to the Patient ID t...

Page 35: ...um by following these steps 1 Plug a USB memory drive minimum 512K in to the USB port on the side of the monitor 2 From the main menu press the SETUP key and then select VOL CLOCK USB DRIVE key 3 Sele...

Page 36: ...rd you until it is completely open 2 Reach in and remove the spent paper core by pulling it gently toward you 3 Place a new paper roll between the two round tabs of the paper holder 4 Pull some paper...

Page 37: ...llowing procedure to select the printing mode to be used Selections are DIRECT TIMED DELAY HR VAR and X RAY The print mode is indicated in the left center of the display 1 Press the SETUP key from the...

Page 38: ...m s 20 seconds before and 20 seconds after at 25mm s HR VAR This mode enables an additional delay mode where the printout is triggered by a beat to beat heart rate change of a given percentage 10 50 i...

Page 39: ...ss key to reset all alarms except LEAD OFF HR HIGH The high heart rate alarm limit has been exceeded for four seconds HR LOW The low heart rate alarm limit has been exceeded for four seconds ASYSTOLE...

Page 40: ...mental regulations In the event that internal adjustment or recalibration is necessary refer to the Operation and Service Manual for this equipment Note If no display is visible on the monitor the mon...

Page 41: ...he ON switch is pressed x Trigger pulse is not functional 9 The Auxiliary port connector is plugged into the monitor 9 ECG size is optimal select Lead II x Erratic ECG waveform Heart Rate is not count...

Page 42: ...s rough surface materials or bring them into contact with anything that could scratch the display or the covers x Do not use petroleum based or acetones solutions or other harsh solvents to clean the...

Page 43: ...case of 100 rolls 590386 E8007RR USB memory stick 590297 E8007RK Roll stand To order accessories please contact customer service x Tele 800 247 4614 ext 106 x Tele 203 481 4183 ext 106 x Fax 203 481 8...

Page 44: ...ent Each lead 100 nA dc maximum Electrode Offset Potential r0 5 V DC Lead Off sensing current 56nA Noise 20 PV peak to peak referred to the input with all leads connected through 51 k 47 nF to ground...

Page 45: ...bpm Display Type Active Matrix TFT Color LCD 640x480 Trace Single ECG trace with freeze function Screen Size 13 25cm x 9 94cm 16 5cm 6 5in diagonal Sweep Speed 25 50 mm s Aspect ratio 0 4 standard Use...

Page 46: ...se width 100ms Pulse amplitude 0 to 5V Output Impedance 100 Sensitivity and Threshold Adjustment Fully Automatic Real Time Clock Resolution 1 minute Display 24 hours Power Requirement The real time cl...

Page 47: ...less than or equal to MCV X indicates hazardous substance concentration higher than MCV The data above represents best information available at the time of publication Some consumable or OEM items may...

Page 48: ...the roll stand Pull down the safety pin and slide the monitor on the adapter plate Release the safety pin and make sure safety pin is engaged in the adapter plate The adapter plate has a hole to allo...

Page 49: ...This page is intentionally left blank...