5.3 Program structure

The furnace offers two types of programs:

a. Ivoclar Vivadent programs for Ivoclar

Vivadent materials

b. Individual programs

All programs are equivalent and therefore full-
fledged programs. In each program, all the
parameters can be adjusted.

a) Ivoclar Vivadent programs for Ivoclar

Vivadent materials (see chapter 10.1)

Refer to enclosed Program table

b) Individual programs

Free, individually adjustable programs

When the furnace is delivered ex
works, the Ivoclar Vivadent programs
already contain the recommended
material parameter settings.

However, the parameters can be changed and
overwritten at any time, if required, if the
programs are to be used for other purposes.
Therefore, the 50 programs are also available as
individual programs.

The programs are designed in such a way that
they can be either used as conventional, one-stage
programs or as two-stage programs, if required.
The mode can be changed via the symbol (one- or
two-stage program) by using the + or – key.

5.4 Adjustable parameters and possible value ranges

Symbol

Parameter

Value range

Value range

P

Program number P

1–300

B

Stand-by temperature

100–700 °C

212–1292 °F

S

Closing time (min : sec)

00:18–30:00

t

À

Temperature increase rate

30–140 °C/min

54–252 °F/min

T

Holding temperature

100–1200 °C

212–2192 °F

H

Holding time (min : sec)

00:01–60:00

V1

Vacuum on

0 or 1–1200 °C

0 or 34–2192 °F

V2

Vacuum off

0 or 1–1200 °C

0 or 34–2192 °F

t

À

Temperature increase rate

t2

À

2

nd

stage

30–140 °C/min

54–252 °F/min

T

Holding temperature

2

nd

stage

100–1200 °C

212–2192 °F

H

Holding time

2

nd

stage (min : sec)

00.01–60:00

V1 (V1 2)

Vacuum on

2

nd

stage

0 or 1–1200 °C

0 or 34–2192 °F

V2 (V2 2)

Vacuum off

2

nd

stage

0 or 1–1200 °C

0 or 34–2192 °F

L

Long-term cooling

0 or 50–1200 °C

0 or 122–2192 °F

t L

Cooling temperature rate

0 or 1–50

0 or 32–90

Automatic plausibility check

The furnace is equipped with an automatic plausibility check function. The parameters
(e.g. T 960 but L 1000) are checked upon each program start. In case of contradictory
parameter combinations, the program stops automatically and the respective error
number is indicated.

Summary of Contents for Programat P300/G2

Page 1: ...Operating Instructions Programat P300 G2 s Valid as of software Version 5 0...

Page 2: ...2...

Page 3: ...ey functions 5 3 Program structure 5 4 Adjustable parameters and possible value ranges 5 5 Settings test programs and information 5 6 Explanation of the symbols on the display 5 7 Explanation of the b...

Page 4: ...use 28 Heater plug 29 Heater plug socket 30 Thermocouple plug socket 32 Leaf spring 33 Air vents base 34 Cooling tray 35 Screw for cooling tray 36 Hood 37 Knurled screw for hood 38 Air vents furnace h...

Page 5: ...5 40 25 34 6 10 35 23 47 33 1 7 5 2 3 4 8 9...

Page 6: ...6 49 26 30 29 50 51 27 28 11 32 43 37 39 36 11 38 25 19 53 13 12 12 13 15 19 46 21 17 18 16...

Page 7: ...losing time 83 Temperature increase 84 Holding temperature 85 Holding time 86 Vacuum on 87 Vacuum off 88 Long term cooling 89 Power saving key 90 Open furnace head 91 Close furnace head 92 Numeric key...

Page 8: ...cation Burn hazard Risk of crushing The Operating Instructions must be read 1 3 Notes regarding the Operating Instructions Furnace concerned Programat P300 G2 Target group Dental technologists These O...

Page 9: ...e must be observed The furnace must be operated under the indicated environmental and operating conditions Chapter 9 The Programat P300 G2 must be properly maintained 2 1 1 Risks and dangers The furna...

Page 10: ...l be prematurely damaged 2 1 11 Risk of crushing and burn hazard Never reach under the furnace head with the hand or other parts of the body during operation since there is a risk of crushing and a bu...

Page 11: ...original spare parts The temperature range for faultless operation is 5 C to 40 C 41 F to 104 F If the furnace has been stored at very low temperatures or high atmospheric humidity the head has to be...

Page 12: ...hat a vacuum may be created with a vacuum pump The firing process is controlled with the corresponding electronic con trols and a software Moreover the set and actual temperatures are continuously com...

Page 13: ...mply use the corresponding inserts Fold the side flaps 2 and combine the two packaging parts by means of the transportation flaps The packaging may be disposed of with the regular household refuse 4 2...

Page 14: ...5 on the firing plate holder 48 Secure the cooling tray 34 with the two screws 35 including the silicone washer 47 Place the cooling tray 34 on the frame plate 7 Make sure that the cooling tray 34 is...

Page 15: ...ace head mounting 43 Ensure that the furnace head mounting mark 41 is aligned with the furnace base mounting mark 42 Make sure that the firing plate 5 is not damaged by mounting the furnace head Step...

Page 16: ...ng additional connections Power connection Please make sure that the voltage indicated on the rating plate complies with the local power supply Connect the power cord 16 with the power socket 17 of th...

Page 17: ...ion I The furnace will now automatically conduct a self test The performance of all furnace components is automatically checked The display shows the following indications during the self test 1 Statu...

Page 18: ...ix non blinking frame around the value Enter keys 76 77 92 The values can be entered by using the keys or the numeric keys Each value entered by means of the or key will be immediately accepted as lon...

Page 19: ...ne or two stage program by using the or key 5 4 Adjustable parameters and possible value ranges Symbol Parameter Value range Value range P Program number P 1 300 B Stand by temperature 100 700 C 212 1...

Page 20: ...ntered by using the numeric keys The date can be selected by using the numeric keys Settings Write protection Ivoclar Vivadent optimized temperature control function Power saving mode Service interval...

Page 21: ...user code The firing hour count can be reset to 0 Only after entry of the user code The operating hour count of the vacuum pump can be reset to 0 Only after entry of the user code Settings Serial num...

Page 22: ...lf test is completed To inform the user that the automatic self test has been successfully completed the selected melody is played 2 Furnace head open and temperature below 550 C 1022 F To inform the...

Page 23: ...being displayed Special variant Conducting a one stage program in which only part of the holding time is fired with a vacuum requirement holding temperature 1 holding temperature 2 6 1 3 Operating ind...

Page 24: ...of the furnace head If the cursor is on the spot in the parameter list shown in the figure you can toggle between the following functions by using the or key Thermo Shock Protection Pre drying no func...

Page 25: ...s and do not influence the function of the furnace in a negative fashion weekly monthly weekly When Thermocouple 4 Check if the thermocouple is straight and in the right place Opening mechanism Check...

Page 26: ...s menu by pressing the Information Settings key Subsequently press and hold the right cursor key 79 until the following display appears To select the calibration pro gram press the ENTER key The follo...

Page 27: ...to start a firing program firing program 27 no Furnace head cannot The furnace head cannot be moved to the final position It might be blocked by an external mechanical be initialized source If not co...

Page 28: ...e following error messages is being displayed 25 29 43 44 45 46 47 48 54 56 103 107 143 144 145 146 147 148 700 701 703 704 705 706 707 1010 1011 1012 1013 1014 1015 1016 1024 1025 1026 1028 1143 1144...

Page 29: ...program still running Wait until the program is complete Switch furnace off and on again If it still does not work properly contact your local Ivoclar Vivadent Service Centre Vacuum pump does not sta...

Page 30: ...he closed furnace Depth 430 mm width 305 mm 410 mm with Cooling tray Height 320 mm Usable size of the firing chamber Diameter 80 mm Height 48 mm Max firing temperature 1200 C 2192 F Weight Furnace bas...

Page 31: ...rvice Centre Important information The current program table is also available at www ivoclarvivadent com downloadcenter The program tables can be downloaded from the internet as PDF files Please make...

Page 32: ...32 10 2 Menu structure 10 2 1 Possibilities of the program selection...

Page 33: ...33 10 2 2 Information Settings...

Page 34: ...34 10 3 Firing curves 10 3 1 Vacuum during both stages 10 3 2 Vacuum during long term cooling 10 3 3 Pre drying 10 3 4 Controlled long term cooling...

Page 35: ......

Page 36: ...7 www ivoclarvivadent com Ivoclar Vivadent S A de C V Av Mazatl n No 61 Piso 2 Col Condesa 06170 M xico D F Mexico Tel 52 55 5062 1000 Fax 52 55 5062 1029 www ivoclarvivadent com mx Ivoclar Vivadent L...

Reviews: