17
4.4 Removing the furnace head
Before the hood (36) is removed, the furnace has
to be switched off and the power cord
disconnected from the power socket (17).
1. Loosen and remove the knurled screw (37) of
the hood (36)
2. Remove the hood (36)
3. Disconnect the thermocouple plug (26)
4. Disconnect the heater plug (28)
5. Press the leaf spring (32) with a finger, lift off
the furnace head at the same time and
remove it
Make sure the furnace head has
completely cooled down before it is
removed (fire hazard).