Configured Options
21 / 120
5001FAOPT71 ISS 8.0
Clock Set
The internal clock is used to record patient history.
1. Turn key switch to
SET
position while
pressing down
START
button.
2. Enter access code using the rate selection
arrows.
3. Enter the first digit using the selection
arrows and press
NEXT
indicated on the
screen.
4. When the complete code has been
entered press
ENTER
.
5. Select
CLOCK SET
from Menu. Press
ENTER
.
6. Display will show current date and time
programmed into unit.
7.
UP/DOWN ARROWS
Use "+" and "-" arrows to change values in
the highlighted field. When entry is correct
press
NEXT
to move highlight from left to
right.
8. Press
OK
to save the date and time.
Important:
The internal clock is the reference against which the PCAM unit stores patient
history and events. Changing the clock will automatically reset the dates against
which all new patient history is stored in the unit and may effect the presentation
of the history graphs
Patient history should always be recorded and, if required, printed prior to
changing the clock.