Exercises for each function
-
ZERO TRIM
- Executing the menu to put ZERO+SPAN button.
- Moving to the sub directory to put Span button when 1 TRIM message appear.
- Executing Zero Trim Function to put the Span button when 11 Z-TRIM message appear.
-
Zero Adjustment : To change the PV value as 14
- Executing the menu to put ZERO+SPAN button.
- Moving to the sub directory to put Span button when 1 TRIM message appear.
- Moving to the sub directory to put Zero button when 11 Z-TRIM message appear.
- Executing Zero Adjustment function to put Span button when 12 Z-ADJ message appear.
- When SelInc message appear, put the Zero button repeatedly until appearing 10.0 message on
LCD. And then, set the value to put Span button.
- When VALUE message appear, change the LCD value 10.0 as to put Zero button, and then put
Zero+Span button.
- When SelInc message appear, change the LCD value 1.0 as to put Zero button, and then set
the value to put Span button.
Put the Zero+Span button after change the LCD value 14.0 when VALUE message appear.
- To save the setting value to put Zero+Span button when SelInc message appear.
-
Change Unit
- Executing the menu to put Zero+Span button.
- Moving to next menu to put Zero button when 1 TRIM message appear.
- Moving to sub directory to put Span button when 2 SETUP message appear.
- Executing the Change Unit function to put Span button when 21 UNIT message appear.
- Saving the setting value to put Span button when the wanted value is displayed on the botton
of LCD to put Zero button repeatedly.
www.
.com