Safety guidelines

After   factory   run-in   and   testing   the   pump   is   drained   of   oil   to   prevent   spillages   during 
shipment - Do not operate the pump until filled with oil. 

Do not use the pump to evacuate combustible, explosive or poisonous gases.

Do not use the pump to evacuate gases that contain particles, dust or steam.

The temperature of the evacuated gas should not exceed 80°C.

Do not use the pump to transfer liquid or to compress gases.

Do not block the air outlet when the pump is operating.

Do not operate the pump in the rain or any other wet environment

.

Oil filling

Ensure that the correct amount of oil is used to fill the pump (see specifications).   Only ITE 
vacuum pump oil should be used as any other oil type may affect performance and could damage 
your pump

.

1.

Remove the exhaust shroud from top front of pump and the cap from the intake fitting.

2.

Pour 50 ml of oil into exhaust hole and then start the pump for a few seconds with the intake 
sealed.

3.

Stop the pump and continue filling until the oil level comes 3/4 of the way up the oil level 
window located on the front of the pump.  DO NOT OVERFILL.

4. Please note that the oil level should be checked only when the pump is running.   When the 

pump is stopped the oil level of a correctly filled pump will fall below the 1/2 way level in the 
sight glass.

Oil change

Contaminants pulled from a system can affect vacuum pump performance and seriously damage 
internal pump parts so therefore frequent oil changes are recommended.  The pump should be hot 
when draining oil and if the pump is not to be used for a long period of time always drain and refill 
the pump with fresh oil before using.

1. Remove the drain plug at the bottom of the front end of the pump in order to drain the used 

oil into a suitable receptacle with pump running.

2.

Refitt the drain plug and refill the pump following the OIL FILLING procedure above.

Gas ballast

The gas ballast valve is used to help prevent condensation of contaminant vapors within the pump. 
Opening the valve stem (knurled rod located on top of pump) allows relatively dry air from the 
atmosphere to enter the pump.

This air combines with the wet vapor passing through the pump and helps prevent the moisture 
from condensing and mixing with the vacuum pump oil.  To use the gas ballast valve slightly open 
the valve during the first few minutes of system pull-down and then close fully to allow the pump 
to achieve ultimate vacuum.

ITE- Sint-Niklaas – Belgium - Fax : +32 3 766 10 67 - [email protected]

Summary of Contents for MK-40-DS

Page 1: ...ITE Vacuum pumps Instruction manual ITE Sint Niklaas Belgium Fax 32 3 766 10 67 sales ite tools com ...

Page 2: ...Table of contents Safety guidelines 3 Oil filling 3 Oil change 3 Gas ballast 3 General operation 4 Parts 5 Specifications 6 Warranty 7 ITE Sint Niklaas Belgium Fax 32 3 766 10 67 sales ite tools com ...

Page 3: ...ould be checked only when the pump is running When the pump is stopped the oil level of a correctly filled pump will fall below the 1 2 way level in the sight glass Oil change Contaminants pulled from a system can affect vacuum pump performance and seriously damage internal pump parts so therefore frequent oil changes are recommended The pump should be hot when draining oil and if the pump is not ...

Page 4: ...uge set or to the system to be evacuated 7 The electric motor has thermal and electric overload protection to prevent the windings from damage if the electric motor becomes too hot or if the current gets too high due to seizer or other mechanical damage 8 The pump is dual cycle and will operate on both 50 Hz and 60 Hz The minimum cold start temperature of the pump is 0 C and below this temperature...

Page 5: ...5 Parts ITE Sint Niklaas Belgium Fax 32 3 766 10 67 sales ite tools com ...

Page 6: ...imensions 330 x 135 x 255 mm 350 x 143 x 265 mm Net weight 9 80 kg 11 50 kg Model MK 120 DS MK 180 DS Voltage 220 V 50Hz 110 V 60Hz 220 V 50 Hz 110 V 60 Hz Free air displacement 132 L min 156 L min 176 L min 212 L min 4 66 CFM 5 50 CFM 6 23 CFM 7 50 CFM Ultimate vacuum 15 micron 15 micron 12 micron 12 micron 0 020 mBar 0 020 mBar 0 016 mBar 0 016 mBar Motor 0 25 kW 0 37 kW 0 37 kW 0 50 kW 1 3 HP 1...

Page 7: ...ps also available with special seals and steel inlet Warranty This ITE vacuum pump manufactured in China to an exclusive ITE product design is guaranteed for 1 year against defects in materials or workmanship when used in accordance with our instructions This warranty is limited to the repair or replacement at our option of products which in our opinion are defective In no circumstances will we al...

Reviews: