background image

PROBLEMS

CAUSE

Sensor pasivado por 

trabajar a más de 3 ppm

El pH del agua es más 

elevado que en el momento 

de la calibración, y no hay 

medida de pH en el 

controlador .

El pH es superior a 9, y por 

lo tanto está fuera de la 

zona de compensación.

SOLUTION

- Check connections

- Realizar una limpieza 

electroquímica

- Calibrar el sensor 

introduciendo el valor 

correcto de pH, o instalar 

un sensor de pH para 

efectuar la compensación 

automática.

-Ajustar el pH dentro del 

margen de pH admisible: 

6.5-9

7.4 PROBLEM-CAUSE-SOLUTION

READING mg/l = 0,

DOES NOT MATCH

WITH DPD-1

MEASUREMENT

Sensor fails to connect with 

control device

Insufficient flow in sensor 

holder, or the chlorine sensor 

not in contact with the water

- Adjust the flow that 

reaches the sensor 

holder. Clean the sensor

holder's filter and flow

regulator.

- Purge the sensor holder 

and ensure that no air is 

left in the measurement 

zone.

- Let the water containing

free chlorine flow around 

the sensor holder for 1 

hour.

- Adjust the flow that 

reaches the sensor 

holder. Clean the sensor

holder's fliter and flow 

regulator.

- Purge the sensor holder 

and ensure there is no air 

in the measurement zone.

Air bubbles in the sensor

measurement zone.

The sensor has been 

measuring water without free 

chlorine for a few hours

READING LOWER

THAN THE DPD-1

MESURE

Insufficient flow in sensor

holder

Air bubbles in sensor

measurement zone.

12

Summary of Contents for 44-010

Page 1: ...ESPA OL Cl2 SENSOR...

Page 2: ......

Page 3: ...ts 10 7 3 Electrode head replace 11 7 4 Problem Cause Solution 12 CE DECLARATION OF CONFORMITY 15 WARRANTY 15 SAFETY RULES In order to prevent personal risks and damages to the environment and to ensu...

Page 4: ...be used in pressure applications and with solids in suspension Manufactured from materials that ensure perfect operation in applications such as Drinking water desininfection Industrial processes Coo...

Page 5: ...Ag AgCl Counter electrode Au GND Electrode Au Scale reading 0 02 3 00 mg l Precision 2 WorKing conditions pH 6 5 9 0 Temperature 0 40 C Salinity 500 ppm Cl 500 ppm SO4 2 Conductivity 50 3000 uS cm Ma...

Page 6: ...d in order to keep the signal as stable as possible This electrodes also made of gold acts as a ground connection to eliminate any residual current that might be found in the water It is important to...

Page 7: ...2 4 3 Working electrode Reference electrode GND Counter electrode 1 2 3 4 The sensor must be installed where it is possible to ensure a constant flow of water with no chance of air bubbles forming in...

Page 8: ...after 24 hours First point calibration 0 mg l Once the sensor is properly conditioned let the water flow at 0 ppm until a stable reading is achieved To facilitate the calibration at 0 mg l the user s...

Page 9: ...ystem wait 60 minutes for the correct polarization of the sensor Calibrate and after 24 hours repeat the calibration If the sensor has been working at 0 mg l without water flow or without water for 1...

Page 10: ...E 44 015 Chlorine sensor electrode head 44 016 Chlorine sensor body 44104 Ring holder for electrodes base 44106 Hydrodynamic regulator 62300 O Ring 19x3 FPM 64349 O Ring 14x2 5 FPM 64410 O Ring 19x2 F...

Page 11: ...plastic tool to prevent electrodes damage Do not damage the connection pins of the electrodes head It is necessary to carry out the step 3 before the assembly otherwise the connection pins of the elec...

Page 12: ...connect with control device Insufficient flow in sensor holder or the chlorine sensor not in contact with the water Adjust the flow that reaches the sensor holder Clean the sensor holder s filter and...

Page 13: ...inestable y el regulador de caudal no act a Hay burbujas de aire en la zona de medici n del sensor Repetir acondicionamiento del sensor y volver a calibrar Repetir medida DPD 1 con reactivos nuevos I...

Page 14: ...nestable y no hay medida de pH en el controlador Eliminar la fuente de la perturbaci n Puede ser til conectar el agua con una toma tierra No utilizar m s de un oxidante para la desinfecci n el agua Es...

Page 15: ...xcepto las partes que sufren desgaste como v lvulas juntas conexiones mangueras y filtro contra todo defecto de fabricaci n o material siempre que la instalaci n uso y mantenimiento del equipo hayan s...

Page 16: ...Ed 03 07 19 ES Manual original P O Box 60 08130 Santa Perp tua de Mogoda BARCELONA Tel 93 544 30 40 Fax 93 544 31 61 C Vall s 26 Pol Ind Can Bernades Subir e mail itc itc es www itc dosing pumps com...

Reviews: