WatchPAT
™
ONE System
38
Operation Manual
Error Message
Possible Reason
Action
registered
Please plug your phone in a charger
No charger plugged into a
mobile phone
Plug in a charger
Data from WatchPAT ONE device finished
transferring. Please open the application to
upload data to your doctor.
The Application could have
been suspended by your
Phone before completion of
the data upload.
Open the WatchPAT ONE
Application and follow the guidance
provided in its screens.
Attention: The WatchPAT device cannot be
communicated. Please bring it closer to the
Application.
The device is not in proximity
or the battery has been
extracted
Bring the device closer to the phone
or insert the battery
6.2
Device Error Messages
If an error occurs and the LED on the device’s cover is blinking, you should take the actions
specified below. If the problem persists you may contact Itamar or an authorized
representative directly.
Device LED’s status
Possible Reason
Action
No light
Battery depleted, or battery
is placed backward or
hardware error.
Check the battery. If it is good and
placed properly there is a hardware
error. Return device to Itamar.
Red LED blinks (every 10 seconds)
The Mobile phone ran out of
power
Connect a charger, rerun the
Application and keep the device close
by until all the stored data has been
sent to the Application
Red LED blinks (every 10 seconds)
The Mobile phone and the
device are not close enough
Make sure the phone and the device
are nearby, until all the stored data
has been sent to the Application
Red LED blinks fast (5 times per
second)
There is a hardware failure in
the device
Return the device to the provider,
and a new one will be shipped in
return.
Red LED is on
The device battery is running
out of power
Replace the battery with a fresh
Alkaline AAA battery