The following table is shown correlation conductivity with concentration of KCl
solution.
K Cl solution(M )
Conductivity
0.001
146.9
S/cm
0.01
1413.0
S/cm
0.05
6.67 mS/cm
0.1
12.89 mS/cm
1
111.9 mS/cm
Compensation Temperature
After setting standard solution, press
Setup
key.
Press
or
key to change 25.0 or 20.0.
The conductivity of a solution exhibits at 25.0
C or 20.0
C.
Data Log Setup
Set by using
or
key
When this setup is set ¡ on¡ , meter transmits data with time
interval of one second.
This data can be taken by communication program or printer
via RS232C interface cable.
When conductivity is 1410
S and temperature is 25.0
C,
the following figure is an example to print.
Chapter IV Setup Functions
N eo M et
SETUP
25.0
COND 1413uS 25.0
SETUP
DATA
oFF