Page 5 of 24
Pegasus R Iss.07 – 08/13
EMC INFORMATION
This product meets the requirements of the European Directive on Electromagnetic Compatibility (EMC)
2004/108/EC and the European Low Voltage Directive BS EN 61010-1
Amendment 1
. To ensure emission
compliance please ensure that any serial communications connecting leads are fully screened.
The product meets the susceptibility requirements of EN 50082-1, criterion B.
Symbol Identification
Publication
Description
ISO3864
Caution (refer to manual)
IEC 417
Caution, Hot Surface
ELECTRICAL SAFETY
This equipment must be correctly earthed.
This equipment is a Class 1 Appliance. A protective earth is used to ensure the conductive parts cannot become live
in the event of a failure of the insulation.
The protective conductor of the flexible mains cable which is coloured green/yellow MUST be connected to a suitable
earth.
The blue conductor should be connected to Neutral and the Brown conductor to Live (Line).
Warning: Internal mains voltage hazard. Do not remove the panels.
There are no user serviceable parts inside. Contact your nearest Isotech agent for repair.
Voltage transients on the supply must not exceed 2.5kV.
Conductive pollution e.g. Carbon dust, must be excluded from the apparatus. EN61010 pollution degrees 2
The apparatus has two input connectors for temperature sensors; see Figure 1. These inputs are only suitable for
either a thermocouple or resistance thermometer. No other sensor or signal may be connected.
Environmental Ratings
Operating Temperature
0-50°C
Relative Humidity
5-95%, non condensing
Insert Warning
The inserts are specially processed for the use with the Pegasus. It is important that only inserts supplied by
Isothermal Technology Ltd are used. Failure to comply with this information may result in damage which would not
be covered under warranty.